R.C.S.A. § 19-25d-10
Special requirements for x-ray therapy equipment operated at potentials of sixty kv and below
Cite as Conn. Agencies Regs. § 19-25d-10
and below
(a) Equipment
(1) All provisions of section 19-25d-9 (a) apply except, for equipment used for "contact
therapy," subdivision (1) in which instance the leakage radiation at the surface of
the tube housing shall not exceed one-tenth roentgen per hour.
(2) There shall be on the control panel some easily discernible device which will give
positive information as to whether or not the tube is energized.
(b) Operating procedures
(1) Automatic timers shall be provided which will permit accurate presetting and determination
of exposures as short as one second.
(2) In the therapeutic application of apparatus constructed with beryllium or other low-filtration
windows, the owner shall insure that the unfiltered radiation reaches only the part
intended and that the useful beam is blocked at all times except when actually being
used.
(3) Machines having an output of more than one thousand roentgens per minute at any accessible
place shall not be left unattended without the power being shut off at the primary
disconnecting means.
(4) If the tube is hand-held during irradiation, the operator shall wear protective gloves
and aprons.