State Operations Manual (Pub. 100-07), Ch. 6 § 6130.4
Choosing the Appropriate Condition
6130.4 - Choosing the Appropriate Condition
(Rev. 199, Issued: 01-17-20, Effective: 01-17-20, Implementation: 01-17-20)
Review the regulatory language at each of the conditions, noting the requirements that
must be met for the condition to be in compliance. For example: The condition of
Bacteriology at 42 CFR §493.1201 (D5002) states the laboratory must meet the
requirements at 42 CFR §§493.1230 (D5200) through 493.1256 (D5485), 493.1261
(D5501 - D5507) and 493.1281 (D5775) through 493.1299 (D5893) (covering General
Laboratory Systems, Preanalytic Systems, Analytic Systems, and Postanalytic
Systems). Serious problems in one or more of these areas can cause the condition of
Bacteriology to be out of compliance.
In comparison, the condition statement for Preanalytic Systems at 42 CFR §493.1240
(D5300) states the laboratory must meet the requirements at 42 CFR §§493.1241
(D5301 - D5309), 493.1242 (D5311 - D5317), and 493.1249 (D5393) for each specialty
or subspecialty of testing. Serious preanalytic deficiencies that are pervasive throughout
the laboratory (not related to specific specialties or subspecialties) could cause the
condition of Preanalytic Systems to be out of compliance. Caution: An enforcement
action based on noncompliance with the condition of General Laboratory Systems,
Preanalytic Systems, Analytic Systems or Postanalytic Systems could be a revocation or
a suspension of the CLIA certificate and would not necessarily be a limitation of the
CLIA certificate for one or more specialties.
Standard-level deficiencies written in one subpart cannot be the basis for a condition in
another subpart. Deficiencies in Proficiency Testing or Personnel would not be the basis
for the condition of Bacteriology to be out of compliance. It is not uncommon for a
surveyor to identify issues that crossover between subparts of the regulations. Cite
deficiencies at the appropriate area of the regulations that describes the problem. For
example, failures in proficiency testing may be caused by an error in specimen
identification, test system malfunction, or lack of training for staff. Consider citing the
most appropriate citation for the laboratory to come into compliance. Avoid citing
multiple citations for the same deficiency unless each citation focuses on a different
aspect of the deficiency (instrument malfunction vs. staff training, or quality system vs.
laboratory director responsibilities, as discussed above).
The surveyor must consider the deficiencies cited when determining the conditions out
of compliance, and also the potential enforcement actions should the laboratory not
correct the deficiencies. The organization of the regulations and conditions allows the
surveyor to write a condition out of compliance according to specialty/subspecialty or to
the Systems of testing (General Laboratory Systems, Preanalytic Systems, Analytic
Systems, or Postanalytic Systems).
EXAMPLE 1:
A laboratory has one or more standard-level deficiencies related to Bacteriology testing
in Preanalytic Systems 42 CFR §493.1241 through 493.1249 (D5301-D5393), Quality
Control Procedures 42 CFR § 493.1256 (D5441-D5485) and the Bacteriology
subspecialty
42 CFR § 493.1261 (D5501-D5507). The surveyor may determine the condition of
Bacteriology 42 CFR § 493.1201 (D5002) is out of compliance based on the deficiencies
written under all three systems, Preanalytic, Analytic and Postanalytic. Even though the
laboratory conducts testing in other specialty or subspecialty areas, by citing the
deficiencies under the condition of Bacteriology, the certificate could be limited for the
subspecialty of Bacteriology instead of the entire CLIA certificate being affected.
EXAMPLE 2:
A laboratory is cited for one or more standard-level deficiencies in Preanalytic Systems
42 CFR §493.1241 through 493.1249 (D5301-D5393) and the deficiencies are related
to practices in all the specialties and subspecialties offered by the laboratory. The
surveyor determines the condition of Preanalytic Systems is out of compliance. If the
laboratory does not correct the condition-level deficiency in Preanalytic Systems, the
enforcement action is against the certificate and not a limitation of a specialty or
subspecialty.
EXAMPLE 3:
A laboratory has deficiencies in Bacteriology in the Control Procedures 42 CFR
§493.1256 (D5441-D5485), the Bacteriology subspecialty 42 CFR §493.1261 (D5501-
D5507), and Routine Chemistry deficiencies in the Control Procedures 42 CFR
§493.1256 (D5441- D5485). All deficiencies are within the Analytic System.
The surveyor may determine the condition of Bacteriology 42 CFR §493.1201 (D5002)
is out of compliance based on the deficiencies cited in Control Procedures 42 CFR
§493.1256 (D5441-D5485) and also deficiencies in subspecialty areas for Bacteriology
42 CFR §493.1261 (D5501-D5507).
And the surveyor may determine the condition of Routine Chemistry 42 CFR
§493.1210 (D5016) is out of compliance based on deficiencies cited related to Control
Procedures 42 CFR §493.1256 (D5441-D5485). Even though the D-tags used to
determine condition- level noncompliance in Routine Chemistry are cited in the
Control Procedures area, the appropriate condition to mark out of compliance is the
applicable subspecialty of Routine Chemistry.
If the laboratory performs testing in only the subspecialties of Bacteriology and
Routine Chemistry, and if the deficient practices are pervasive, the surveyor may
write the condition of Analytic Systems 42 CFR §493.1250 (D5400) out of
compliance.
When a specialty or subspecialty condition is out of compliance, the enforcement action
chosen may be a limitation to the certificate for the specialty or subspecialty out of
compliance. This approach allows the laboratory to continue testing in those specialties
and subspecialties in which compliance was determined. A condition-level deficiency in
one of the Systems (General Laboratory Systems, Preanalytic Systems, Analytic
Systems, or Postanalytic Systems) indicates a pervasive situation through all specialties
and subspecialties offered by the laboratory.