40 C.F.R. § 1065.366
Interference verification for FTIR analyzers.
Cite as 40 C.F.R. § 1065.366 (2026)
(a) Scope and frequency. If you measure CH 4 , C 2 H 6 , NMHC, or NMNEHC using an FTIR analyzer, verify the amount of interference after initial analyzer installation and after major maintenance. (b) Measurement principles. Certain species can interfere with analyzers by causing a response similar to the target analyte. If the analyzer uses compensation algorithms that utilize measurements of other gases to meet this interference verification, a correct result depends on simultaneously conducting these other measurements to test the compensation algorithms during the analyzer interference verification. (c) System requirements. An FTIR analyzer must have combined interference that is within ±2% of the flow-weighted mean concentration of CH 4 , NMHC, or NMNEHC expected at the standard, though we strongly recommend a lower interference that is within ±1%. (d) Procedure. Perform the interference verification for an FTIR analyzer using the same procedure that applies for N 2 O analyzers in § 1065.375(d). [81 FR 74168, Oct. 25, 2016, as amended at 89 FR 29801, Apr. 22, 2024]
- Cross-references to the CFR
- 1065.375