42 C.F.R. § 84.155

Airflow resistance test; Type C supplied-air respirator, continuous flow class and Type CE supplied-air respirator; minimum requirements.

Last amended: 2015Year: 2026Length: 34 wordsOfficial source

Cite as 42 C.F.R. § 84.155 (2026)

The resistance to air flowing from the respirator shall not exceed 25 mm. (1 inch) of water-column height when the air flow into the respiratory-inlet covering is 115 liters (4 cubic feet) per minute.
42 C.F.R. § 84.155: Airflow resistance test; Type C supplied-air respirator, continuous flow class and Type CE supplied-air respirator; minimum requirements. | Justis AI