42 C.F.R. § 84.203
Breathing resistance test; minimum requirements.
Cite as 42 C.F.R. § 84.203 (2026)
(a) Resistance to airflow will be measured in the facepiece, mouthpiece, hood, or helmet of a chemical cartridge respirator mounted on a test fixture with air flowing at a continuous rate of 85 liters per minute, both before and after each test conducted in accordance with §§ 84.206 through 84.207.
(b) The maximum allowable resistance requirements for chemical cartridge respirators are as follows:
Maximum Resistance
[Millimeter water column height]
Type of chemical-cartridge respirator
Inhalation
Exhalation
Initial
Final
1
Other than single-use vinyl chloride respirators:
For gases, vapors, or gases and vapors
40
45
20
For gases, vapors, or gases and vapors, and particulates
50
70
20
Single-use respirator with valves:
For vinyl chloride
20
25
20
For vinyl chloride and particulates
30
45
2
Single-use respirator without valves:
For vinyl chloride
15
20
(
2
)
For vinyl chloride and particulates
25
40
(
2
)
1
Measured at end of service life specified in Table 11 of this subpart.
2
Same as inhalation.