42 C.F.R. § 84.90

Breathing resistance test; inhalation.

Last amended: 2015Year: 2026Length: 76 wordsSubsections: 3Official source

Cite as 42 C.F.R. § 84.90 (2026)

(a) Resistance to inhalation airflow will be measured in the facepiece or mouthpiece while the apparatus is operated by a breathing machine as described in § 84.88. (b) The inhalation resistance of open-circuit apparatus shall not exceed 32 mm. (1.25 inch) water-column height (at a flow rate of 120 liters per minute). (c) The inhalation resistance of closed-circuit apparatus shall not exceed the difference between exhalation resistance (§ 84.91(e)) and 10 cm. (4 inches) water-column height.
Cross-references to the CFR
84.88.84.91