47 C.F.R. § 15.241

Operation in the band 174-216 MHz.

Last amended: 2016Year: 2026Length: 109 wordsSubsections: 3Official source

Cite as 47 C.F.R. § 15.241 (2026)

(a) Operation under the provisions of this section is restricted to biomedical telemetry devices. (b) Emissions from the device shall be confined within a 200 kHz band which shall lie wholly within the frequency range of 174-216 MHz. (c) The field strength of any emissions radiated within the specified 200 kHz band shall not exceed 1500 microvolts/meter at 3 meters. The field strength of emissions radiated on any frequency outside of the specified 200 kHz band shall not exceed 150 microvolts/meter at 3 meters. The emission limits in this paragraph are based on measurement instrumentation employing an average detector. The provisions in § 15.35 for limiting peak emissions apply.
Cross-references to the CFR
15.35
47 C.F.R. § 15.241: Operation in the band 174-216 MHz. | Justis AI