47 C.F.R. § 95.2565

MedRadio frequency accuracy.

Last amended: 2017Year: 2026Length: 71 wordsOfficial source

Cite as 47 C.F.R. § 95.2565 (2026)

Each MedRadio transmitter type must be designed to maintain a frequency stability of ±100 ppm of the operating frequency over the applicable temperature range set forth in this section. Frequency stability testing shall be performed over the appropriate temperature range. (a) 25 °C to 45 °C in the case of medical implant transmitters; and (b) 0 °C to 55 °C in the case of MedRadio programmer/control transmitters and medical body-worn transmitters.