10 C.F.R. § 35.457

Therapy-related computer systems.

Last amended: 2016Year: 2026Length: 87 wordsSubsections: 4Official source

Cite as 10 C.F.R. § 35.457 (2026)

The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of: (a) The source-specific input parameters required by the dose calculation algorithm; (b) The accuracy of dose, dwell time, and treatment time calculations at representative points; (c) The accuracy of isodose plots and graphic displays; and (d) The accuracy of the software used to determine sealed source positions from radiographic images.