64E-5.645. Remote Afterloader, Gamma Stereotactic, and Teletherapy Therapy-Related Computer Systems

Last amended: 2013Year: 2026Length: 136 wordsOfficial source

Cite as Fla. Admin. Code r. 64E-5.645

The licensee shall perform acceptance testing on the treatment planning system of high, medium, low, pulsed dose-rate remote afterloaders, gamma stereotactic, and teletherapy therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. An example of a nationally recognized body is the American Association of Physicists in Medicine. At a minimum, the acceptance testing must include, as applicable, verification of the following: (1) The source-specific input parameters required by the dose calculation algorithm; (2) The accuracy of dose, dwell time, and treatment time calculations at representative points; (3) The accuracy of isodose plots and graphic displays; (4) The accuracy of the software used to determine sealed source positions from radiographic images; and (5) The accuracy of electronic transfer of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.