HAR §11-45-101. Radiation survey instruments
Cite as Haw. Code R. § 11-45-101
(a) The licensee
shall maintain sufficient calibrated and operable radiation survey
instruments to make physical radiation surveys as required by this
subchapter and subchapter 4. Instrumentation required by this
section shall have a range such that two milliroentgens (5.16 x
10-7 coulombs per kilogram) per hour through one roentgen (2.58 x
10-4 coulombs per kilogram) per hour can be measured.
(b)
Each radiation survey instrument shall be calibrated:
(1)
At energies appropriate for use and at intervals not
to exceed three months and after each instrument
servicing;
(2)
Such that accuracy within plus or minus twenty percent
can be demonstrated; and
§11-45-104
45-81
(3)
At two points located approximately one-third and two-
thirds of full-scale on each scale for linear scale
instruments; at midrange of each decade, and at two
points of at least one decade for logarithmic scale
instruments; and at appropriate points for digital
instruments.
(c) Records of these calibrations shall be maintained after
the calibration date for inspection by the department for three
years.
(d) Each radiation survey instrument shall be checked with
a radiation source at the beginning of each day of use and at the
beginning of each work shift to ensure it is operating properly.
[Eff 11/12/99] (Auth: HRS §§321-10, 321-11, 321-71) (Imp: HRS
§§321-1, 321-11(21), 321-71)