HAR §11-45-108
HAR §11-45-108. Personnel monitoring control
Cite as Haw. Code R. § 11-45-108
(a) The licensee
shall not permit any individual to act as a radiographer or as a
radiographer trainee unless, at all times during radiographic
operations, the individual wears a direct reading pocket
dosimeter, an alarm ratemeter, and either a film badge or a
thermoluminescent dosimeter (TLD). Pocket dosimeters shall have a
range from zero to two hundred milliroentgens (5.16 x 10-5
coulombs per kilogram) and shall be recharged daily or at the
start of each shift. Each film badge or TLD shall be assigned to
and worn by only one individual.
(b) Pocket dosimeters shall be read and exposures recorded
at least once daily.
(c) Pocket dosimeters shall be checked for correct response
to radiation at periods not to exceed one year. Acceptable
dosimeters shall read within plus or minus thirty percent of the
true radiation exposure. Records of this check shall be
maintained for inspection by the department for three years.
(d)
Each alarm ratemeter shall:
(1)
Be checked to ensure that the alarm functions properly
§11-45-108
45-84
(sounds) before use at the start of each shift;
(2)
Be set to give an alarm signal at a preset dose rate
of five hundred milliroentgens per hour;
(3)
Require special means to change the preset alarm
function; and
(4)
Be calibrated at periods not to exceed one year for
correct response to radiation. Acceptable ratemeters
shall alarm within plus or minus twenty percent of the
true radiation dose rate. Records of these
calibrations shall be maintained for inspection by the
department for three years.
(e) If an individual's pocket dosimeter is inoperable or
discharged beyond its range, industrial radiographic operations by
that individual shall cease and the individual's film badge or TLD
shall be processed immediately. The individual shall not return
to work with x-ray systems until a determination of the radiation
exposure has been made.
(f) Reports received from the film badge or TLD processor
and records of daily pocket dosimeter readings shall be kept for
inspection by the department for three years.
(g) If a film badge or TLD is lost or damaged, the worker
shall cease work immediately until a replacement film badge or TLD
is provided and the exposure is calculated for the time period
from issuance to loss or damage of the film badge or TLD.
[Eff 11/12/99] (Auth: HRS §§321-10, 321-11, 321-71) (Imp: HRS
§§321-1, 321-11(21), 321-71)