HAR §11-45-111
HAR §11-45-111. Radiation surveys and survey records
Cite as Haw. Code R. § 11-45-111
(a) No
radiographic operation shall be conducted unless calibrated and
operable radiation survey instrumentation is available and used at
each site where radiographic exposures are made.
(b) A physical radiation survey shall be made after each
radiographic exposure to determine that the machine is "off".
§11-45-113
45-85
(c) A survey with a radiation survey instrument shall be
made after each radiographic exposure to determine that x-ray
system is not operational. The area at and surrounding the
collimator of the x-ray system shall be surveyed.
(d) Records shall be kept for the required surveys. Such
records shall be maintained for inspection by the department for
three years. [Eff 11/12/99] (Auth: HRS §§321-10, 321-11, 321-
71) (Imp: HRS §§321-1, 321-11(21), 321-71)
§11-45-112 Documents and records required at temporary
jobsites. Each licensee conducting industrial radiography at a
temporary jobsite shall have the following records available at
that site for inspection by the department for three years:
(1)
Copy of license;
(2)
Operating and emergency procedures;
(3)
Chapter 11-45;
(4)
Survey records pursuant to subchapter 5 and area
survey records pursuant to subchapter 4 for the period
of operation at the site;
(5)
Daily pocket dosimeter records for the period of
operation at the site; and
(6)
The latest instrument calibration records for survey
instruments in use at the site. Acceptable records
include tags or labels which are affixed to the device
or survey meter. [Eff 11/12/99] (Auth: HRS §§321-
10, 321-11, 321-71) (Imp: HRS §§321-1, 321-11(21),
321-71)
§11-45-113 Specific requirements for radiographic personnel
performing industrial radiography. (a) At a jobsite, the
following shall be supplied by the licensee:
(1)
At least one operable, calibrated survey instrument;
(2)
A current whole body personnel monitor (TLD or film
badge) for each individual;
(3)
An operable, calibrated pocket dosimeter with a range
of zero to two hundred milliroentgens (5.16 x 10-5
coulombs per kilogram) for each worker; and
(4)
The appropriate barrier ropes and signs.
(b)
Industrial radiographic operations shall not be
performed if any of the items in subsection (a) are not available
at the jobsite or are inoperable.
(c)
Each licensee shall provide a minimum of two
radiographic personnel when x-ray systems are used at temporary
jobsites. If one of the personnel is a radiographer trainee, the
other shall be a radiographer instructor.
(d)
No individual other than a radiographer or a
radiographer trainee who is under the personal supervision of a
radiographer instructor shall manipulate controls or operate
equipment used in industrial radiographic operations.
(e)
No individual shall act as a radiographer instructor
unless such individual:
(1)
Has met the requirements of subchapter 5;
§11-45-113
45-86
(2)
Has one year of documented experience as a
radiographer; and
(3)
Has been named as a radiographer instructor on the
license issued by the department.
(f)
During an inspection by the department, the department
inspector may terminate an operation if any of the items in
subsection (a) are not available and operable or if the required
number of radiographic personnel are not present. Operations
shall not be resumed until such conditions are met.
[Eff 11/12/99] (Auth: HRS §§321-10, 321-11, 321-71) (Imp: HRS
§§321-1, 321-11(21), 321-71)
§11-45-114 Special requirements and exemptions for cabinet
radiography. (a) Systems for cabinet radiography designed to
allow admittance of individuals shall:
(1)
Comply with all applicable requirements of this
subchapter and subchapter 4. If such a system is a
certified cabinet x-ray system, it shall comply with
all applicable requirements of this subchapter and 21
C.F.R. §1020.40.
(2)
Be evaluated at intervals not to exceed one year to
assure compliance with the applicable requirements of
this subchapter. Records of these evaluations shall
be maintained for inspection by the department for
three years.
(b)
Certified cabinet x-ray systems designed to exclude
individuals from the interior of the cabinet are exempt from the
requirements of this subchapter except that:
(1)
Operating personnel shall be provided with either a
film badge or a thermoluminescent dosimeter, and
reports of the results shall be maintained for
inspection by the department for three years.
(2)
No licensee shall permit any individual to operate a
cabinet x-ray system until such individual has
received a copy of and instruction in the operating
procedures for the unit and has demonstrated compe-
tence in its use. Records which demonstrate
compliance with paragraph (1) shall be maintained for
inspection by the department for three years.
(3)
Tests for proper operation of high radiation area
control devices or alarm systems, where applicable,
shall be conducted, recorded, and maintained in
accordance with section 11-45-104.
(4)
The licensee shall perform an evaluation, at intervals
not to exceed one year, to determine conformance with
subchapter 4. If such a system is a certified cabinet
x-ray system, it shall be evaluated at intervals not
to exceed one year to determine conformance with 21
C.F.R. §1020.40. Records of these evaluations shall
be maintained for inspection by the department for
three years.
(c) Certified cabinet x-ray systems shall be maintained in
compliance with 21 C.F.R. §1020.40 unless prior approval has been
§11-45-116
45-87
granted by the department pursuant to subchapter 1.
[Eff 11/12/99] (Auth: HRS §§321-10, 321-11, 321-71) (Imp: HRS
§§321-1, 321-11(21), 321-71)
SUBCHAPTER 6
X-RAYS IN THE HEALING ARTS