HAR §11-45-139
HAR §11-45-139. Quality control of diagnostic equipment
Cite as Haw. Code R. § 11-45-139
Each
licensee shall establish written quality control procedures for
all diagnostic equipment used for radionuclide studies. As a
minimum, quality control procedures and frequencies shall be those
recommended by equipment manufacturers and procedures which have
been approved by the department. The licensee shall conduct
quality control procedures in accordance with written procedures.
[Eff 11/12/99] (Auth: HRS §§321-10, 321-11, 321-71) (Imp: HRS
§§321-1, 321-11(21), 321-71)
§11-45-140 Possession, use, calibration, and check of dose
calibrators. (a) A medical use licensee authorized to administer
radiopharmaceuticals shall possess a dose calibrator and use it to
measure the amount of activity administered to each patient. In
the case where the ionization type dose calibrator cannot be used
effectively to verify administered activity, the licensee shall
use an alternative method. Any alternative method for the use of
a dose calibrator shall be approved by the department. Any
alternative method shall provide for acceptable verification of
constancy, accuracy, linearity, and geometry dependence as
applicable.
(b)
Each licensee shall establish written quality control
procedures for all dose calibrators used for measuring the amount
of activity administered to a patient. As a minimum, quality
control procedures and frequencies shall be those recommended by
the American National Standards Institute:
(1)
Check each dose calibrator for constancy with a
dedicated check source at the beginning of each day of
use. To satisfy the requirement of this section, the
check shall be done on a frequently used setting with
a sealed source of not less than fifty microcuries
(1.85 megabecquerels) with energies representative of
the radionuclides in clinical use at the facility;
(2)
Test each dose calibrator for accuracy upon
§11-45-140
45-124
installation and at intervals not to exceed twelve
months thereafter by assaying at least two sealed
sources containing different radionuclides, the
activity of which the manufacturer has determined
within five percent of the stated activity, with
minimum activity of fifty microcuries (1.85
megabecquerels) and energies representative of the
radionuclides in clinical use at the facility;
(3)
Test each dose calibrator for linearity upon
installation and at intervals not to exceed three
months thereafter over the range of use between ten
microcuries (three hundred seventy kilobecquerels) and
the highest dosage that shall be assayed; and
(4)
Test each dose calibrator for geometry dependence upon
installation over the range of volumes and volume
configurations for which it shall be used. The
licensee shall keep a record of this test for the
duration of the use of the dose calibrator.
(c)
A licensee shall mathematically correct dosage
readings for any geometry or linearity error that exceeds ten
percent if the dosage is greater than ten microcuries (three
hundred seventy kilobecquerels) and shall repair or replace the
dose calibrator if the accuracy or constancy error exceeds ten
percent.
(d)
A licensee shall also perform checks and tests
required by subsection (b) following adjustment or repair of the
dose calibrator.
(e)
A licensee shall retain a record of each check and
test required by this section. The records required by subsection
(b) shall include:
(1)
For subsection (b)(1), the model and serial number of
the dose calibrator, the identity and calibrated
activity of the radionuclide contained in the check
source, the date of the check, the activity measured,
the instrument settings, and the initials of the
individual who performed the check;
(2)
For subsection (b)(2), the model and serial number of
the dose calibrator, the model and serial number of
each source used and the identity of the radionuclide
contained in the source and its activity, the date of
the test, the results of the test, the instrument
settings, and the signature of the individual who
performed the test;
(3)
For subsection (b)(3), the model and serial number of
the dose calibrator, the calculated activities, the
measured activities, the date of the test, and the
signature of the radiation safety officer; and
(4)
For subsection (b)(4), the model and serial number of
the dose calibrator, the configuration and calibrated
activity of the source measured, the activity of the
source, the activity measured and the instrument
setting for each volume measured, the date of the
test, and the signature of the individual who
performed the test.
§11-45-142
45-125
(f)
A licensee shall use dose calibrator reference and
calibration sources traceable to the National Institute of
Standards and Technology or other standards recognized as being
equivalent by the National Institute of Standards and Technology.
[Eff 11/12/99] (Auth: HRS §§321-10, 321-11, 321-71) (Imp: HRS
§§321-1, 321-11(21), 321-71)