HAR §11-45-177. Personnel requirements
Cite as Haw. Code R. § 11-45-177
(a) No individual
shall be permitted to operate or maintain analytical x-ray
equipment unless such individual has received instruction in and
demonstrated competence as to:
§11-45-177
45-144
(1)
Identification of radiation hazards associated with
the use of the equipment;
(2)
Significance of the various radiation warning, safety
devices, and interlocks incorporated into the
equipment, or the reasons they have not been installed
on certain pieces of equipment and the extra
precautions required in such cases;
(3)
Proper operating procedures for the equipment;
(4)
Recognition of symptoms of an acute localized
exposure; and
(5)
Proper procedures for reporting an actual or suspected
exposure.
(b)
Finger or wrist dosimetric devices shall be provided
to and shall be used by:
(1)
Analytical x-ray equipment workers using systems
having an open-beam configuration and not equipped
with a safety device; and
(2)
Personnel maintaining analytical x-ray equipment if
the maintenance procedures require the presence of a
primary x-ray beam when any local component in the
analytical x-ray system is disassembled or removed.
(c)
Reported dose values shall not be used for the purpose
of determining compliance with subchapter 4 unless evaluated by a
qualified health or medical physicist. [Eff 11/12/99] (Auth:
HRS §§321-10, 321-11, 321-71) (Imp: HRS §§321-1, 321-11(21),
321-71)