HAR §11-45-213
HAR §11-45-213. Radiation survey instruments
Cite as Haw. Code R. § 11-45-213
(a) The licensee
shall maintain sufficient calibrated and operable radiation survey
instruments at each field station to make physical radiation
surveys as required by this subchapter and subchapter 4. Instru-
mentation shall be capable of measuring 0.1 milliroentgen (25.8
nanocoulombs per kilogram) per hour through at least
fifty milliroentgens (12.9 microcoulombs per kilogram) per hour.
Survey instruments acquired before the effective date of this
chapter and capable of measuring 0.1 milliroentgen
(25.8 nanocoulombs per kilogram) per hour through at least
twenty milliroentgens (5.16 microcoulombs per kilogram) per hour
also satisfies this requirement five years after the effective
date of this chapter.
(b)
Each radiation survey instrument shall be calibrated:
(1)
At intervals not to exceed six months and after each
instrument servicing;
(2)
For linear scale instruments, at two points located
approximately one-third and two-thirds of full-scale
on each scale; for logarithmic scale instruments, at
midrange of each decade, and at two points of at least
one decade; and for digital instruments, at
appropriate points; and
(3)
So that accuracy within twenty percent of the true
radiation level can be demonstrated on each scale.
(c)
Calibration records shall be maintained at the
facility for three years. [Eff 11/12/99] (Auth: HRS §§321-10,
321-11, 321-71) (Imp: HRS §§321-1, 321-11(21), 321-71)