HAR §11-45-51
HAR §11-45-51. Reserved
Cite as Haw. Code R. § 11-45-51
§11-45-52 Testing for leakage or contamination of sealed
sources. (a) Each sealed source with a half-life greater than
thirty days and in any form other than gas, shall be tested for
leakage or contamination before initial use and, unless otherwise
authorized by the department, at intervals not to exceed six
months, except that each source designed for the purpose of
emitting alpha particles shall be tested at intervals not to
exceed three months. If, at any other time, there is reason to
suspect that a sealed source might have been damaged or might be
leaking, it shall be tested for leakage before further use. In
the absence of a certificate from a transferor indicating that a
test for leakage has been made within six months before the
transfer, the sealed source shall not be put into use until tested
and the results received.
(1)
Tests for leakage for all sealed sources, except those
manufactured to contain radium, shall be capable of
§11-45-52
45-59
detecting the presence of 0.005 microcuries (one
hundred eighty-five becquerels) of radioactive
material on a test sample. Test samples shall be
taken from the sealed source or from the surfaces of
the container in which the sealed source is stored or
mounted on which one might expect contamination to
accumulate.
(2)
The test for leakage for sources manufactured to
contain radium shall be capable of detecting an
absolute leakage rate of 0.001 microcuries (thirty-
seven becquerels) of radon-222 in a twenty-four hour
period when the collection efficiency for radon-222
and its daughters has been determined with respect to
collection method, volume and time.
(3)
Test samples shall also be taken from the interior
surfaces of the container in which sealed sources of
radium are stored. This test shall be capable of
detecting the presence of 0.005 microcuries (one
hundred eighty-five becquerels) of a radium daughter
which has a half-life greater than four days.
(4)
Notwithstanding the periodic test for leakage
required, any sealed source is exempt from such tests
for leakage when the sealed source contains one
hundred microcuries (3.7 megabecquerels) or less of
beta or gamma emitting material or ten microcuries
(three hundred seventy becquerels) or less of alpha
emitting material.
(b)
Tests for leakage or contamination shall be performed
by persons specifically authorized by the department to perform
such services.
(c)
The following shall be considered evidence that the
sealed source is leaking:
(1)
The presence of 0.005 microcuries (one hundred eighty-
five becquerels) or more of removable contamination on
any test sample. If the test of a sealed source,
other than radium, reveals the presence of 0.005
microcuries (one hundred eighty-five becquerels) or
more of removable contamination, the licensee shall
immediately withdraw the sealed source from use, take
action to prevent the spread of contamination, and
cause the sealed source to be decontaminated and
repaired or to be disposed of in accordance with
subchapter 4.
(2)
Leakage of 0.001 microcuries (thirty-seven becquerels)
of radon-222 per twenty-four hours for sealed sources
manufactured to contain radium. If the test of a
sealed source manufactured to contain radium reveals
the presence of removable contamination resulting from
the decay of 0.005 microcuries (one hundred eighty-
five becquerels) or more of radium-226, the licensee
shall immediately withdraw the sealed source from use,
take action to prevent the spread of contamination,
cause the sealed source to be decontaminated and
repaired or to be disposed of in accordance with
§11-45-52
45-60
subchapter 4.
(d)
Test results shall be kept in units of becquerel or
microcurie and maintained for inspection by the department for
three years.
(e)
Reports of test results for leaking or contaminated
sealed sources shall be made pursuant to section 11-45-96.
[Eff 11/12/99] (Auth: HRS §§321-10, 321-11, 321-71) (Imp: HRS
§§321-1, 321-11(21), 321-71)