32 Ill. Adm. Code 333.40
Calibration Services
Section 333
Section 333.40 Calibration
Services
a) The Department provides the following calibration services
under accreditation by the Conference of Radiation Control Program Directors,
Inc.
1) X-ray Calibrations. 5 beam codes at 32 mR/sec. This service
is available for the Radcal MDH 1015 x-ray monitor only. The available beam
codes are:
BEAM
CODE
HALF-VALUE
LAYER
HOMOGENEITY
COEFFICIENT
L
80
1.81 mm
Aluminum
0.59
L
100
2.82 mm
Aluminum
0.59
M
30
0.36 mm
Aluminum
0.65
M
50
1.03 mm
Aluminum
0.64
M
100
4.96 mm
Aluminum
0.73
2) Gamma Ray Calibrations. 1 beam code. This service is
available only for ion chamber and energy compensated G-M tube instruments.
BEAM CODE
INTENSITIES
Cs-137
.005 - 40 R/hr
b) The Department also provides the following services, which are
not performed under accreditation.
1) Microrem Meter Calibrations. 2 points per scale, gamma
intensity range available to 0.015 mR/hr.
2) Pocket Dosimeter Calibrations. 1 point gamma radiation up to
200 R.
3) Count-Rate Instrument Calibrations. Pulse rate at 2 points
per scale. Alpha probe response is checked at 2 points. Beta-gamma probes are
checked at 1 point. Other checks may be performed, depending on each
instrument and probe combination.
c) The Department may be able to provide additional calibration
services, including but not limited to calibrations using additional beam
codes. Persons desiring any calibration services are urged to contact the
facility by telephone at (217) 786-7221 to determine the availability of
additional services.