32 Ill. Adm. Code 335.2010
Possession, Use and Calibration of Instruments Used to Measure the Activity of Unsealed Radioactive Material
Section 335
Section 335.2010 Possession,
Use and Calibration of Instruments Used to Measure the Activity of Unsealed
Radioactive Material
a) For direct measurements performed in accordance with Section
335.2030, the licensee shall possess and use instrumentation to measure the activity
of unsealed byproduct material before it is administered to each patient or
human research subject.
b) A licensee shall calibrate the instrumentation required in
subsection (a) in accordance with nationally recognized standards or the
manufacturer's instructions.
c) A licensee shall maintain a record of instrument calibrations
required by subsection (b) for 5 years. The records shall include the model
and serial number of the instrument, the date of the calibration, the results
of the calibration, the name of the individual who performed the calibration
and a copy of the national standard or manufacturer's instructions used to
perform the calibration.