32 Ill. Adm. Code 335.2040
Authorization for Calibration, Transmission, Attenuation Correction and Reference Sources
Section 335
Section 335.2040
Authorization for Calibration, Transmission, Attenuation Correction and
Reference Sources
Any person authorized by Section
335.30 for medical use of radioactive material may receive, possess and use the
following radioactive material for check, calibration, transmission,
attenuation correction and reference use. Reference sources containing
radioactive material authorized under this Part shall not be used for medical
use except in accordance with the requirements in Section 335.6010. Sealed
sources shall not be combined (i.e. bundled or aggregated) to create an
activity greater than the maximum activity of any single sealed source
authorized under this Section. Sealed sources are authorized as follows:
a) Sealed sources not exceeding 1.11 GBq (30 mCi) each, manufactured
and distributed by a person licensed under 32 Ill. Adm. Code 330.280(k) or
equivalent U.S. Nuclear Regulatory Commission or Agreement State regulations.
b) Sealed sources, not exceeding 1.11 GBq (30 mCi) each,
redistributed by a licensee authorized to redistribute the sealed sources
manufactured and distributed by a person licensed under 32 Ill. Adm. Code
330.280(k) or equivalent U.S. Nuclear Regulatory Commission or Agreement State
regulations, providing the redistributed sealed sources are in the original
packaging and shielding and are accompanied by the manufacturer's approved
instructions.
c) Any radioactive material with a half-life not greater than 120
days in individual amounts not to exceed 0.56 GBq (15 mCi).
d) Any radioactive material with a half-life longer than 120 days
in individual amounts not to exceed the smaller of 7.4 MBq (200
m
Ci) or 1000 times the quantities in Appendix
B of 32 Ill. Adm. Code 330.
e) Technetium-99m in amounts as needed.