32 Ill. Adm. Code 335.2080
Monitoring for Contamination and Ambient Radiation Dose Rate
Section 335
Section 335.2080 Monitoring
for Contamination and Ambient Radiation Dose Rate
a) In addition to the monitoring required by 32 Ill. Adm. Code
340, the licensee shall measure with a radiation detection survey instrument
all areas where unsealed radioactive material was prepared for use or
administered at the end of use each day. However, the licensee does not need
to perform the monitoring required by this Section in areas where patients or
human research subjects are confined until release under Section 335.2110. The
instrument shall be operable and calibrated in accordance with the requirements
of 32 Ill. Adm. Code 340.510(b) and (c).
b) At least once each week, a licensee shall measure with a
radiation detection survey instrument all areas where radiopharmaceuticals or
radioactive wastes are stored to ensure compliance with 32 Ill. Adm. Code
340.210 and 340.310. The instrument shall be operable and calibrated in
accordance with the requirements of 32 Ill. Adm. Code 340.510(b) and (c).
c) At least once each week, a licensee shall measure for
removable contamination in all areas where unsealed radioactive materials are
prepared for use, administered or stored.
d) A licensee shall conduct the measurements required by subsections
(b) and (c) in a manner that permits detection of both external exposure rates
and removable contamination that would give rise to exposures in excess of the
limits specified in 32 Ill. Adm. Code 340.210 and 340.310.
e) A licensee shall retain a record of each survey required by
this Section for 5 years. The record shall include the monitoring date, a description
of each area monitored, the measurement results, the manufacturer, model and
serial number of the instruments, and the identity of the individual who
performed the monitoring.
AGENCY NOTE: For
the purposes of this Section, 2000 dpm (disintegrations per minute) per 100
square centimeters of surface area may be utilized as a sufficiently sensitive
detection limit for removable contamination unless the licensee has developed
alternate removable contamination limits which take into consideration the
unsealed radionuclides in use, their respective contribution to the dose limits
in 32 Ill. Adm. Code 340.210 and 340.310, and the detection capability of the
radiation detection survey instruments in use. Measurement of removable
contamination shall only be performed with a survey instrument, in lieu of
wipes, if the instrument is sufficiently sensitive to detect the contamination
at the limits specified in this Section.