32 Ill. Adm. Code 335.2150
Additional Technical Requirements for Intravascular Brachytherapy Units
Section 335.2150 Additional Technical Requirements for
Intravascular Brachytherapy Units
In addition to other provisions required by this Part, the
licensee authorized to use an intravascular brachytherapy unit for medical use
shall:
a) Have
a treatment team consisting of, at a minimum, an interventional cardiologist,
an authorized user and an authorized medical physicist and that, at a minimum,
an interventional cardiologist and an authorized user will be physically
present in the treatment suite during all radioactive procedures.
AGENCY NOTE: The requirements of
32 Ill. Adm. Code 401 regarding radiation therapists must also be met.
b) Independently
verify source strength and uniformity. Dwell time at the treatment location
must be monitored and recorded. Source uniformity or strength must not differ
by more that 10 percent of the expected values.
c) For
devices requiring additional shielding, demonstrate compliance with 32 Ill.
Adm. Code 340.210 and 340.310 requirements.
d) Inspect
sealed sources, source trains or ribbons before each use and ensure sources are
removed from service at intervals established by the manufacturer (i.e.,
confirm that source trains will not be used after the "use by" date,
at intervals not to exceed two months from the date of shipment, or when
evidence of degradation is observed, whichever comes first).
e) Inspect
and service devices containing sealed sources at intervals established by the
manufacturer, and ensure that maintenance and repair of the device is performed
only by the manufacturer or persons specifically authorized by the Agency, the
U.S. Nuclear Regulatory Commission, an Agreement State or a Licensing State to
perform such service.
f) Prohibit
cuts, alterations or splicing of the sealed sources, source trains or ribbons,
except in situations involving an emergency where the source wire cannot be
returned to its normal safe position. If such cuts, alterations or splicing are
necessary, notification in accordance with Section 335.1080 or 32 Ill. Adm.
Code 340.1220 shall be made to the Agency.
g) Use
only manufacturer provided inducer sheaths, catheters and accessories to ensure
their demonstrated equivalents will be used with the devices.
h) Ensure
the daily operational checks will be performed prior to patient treatment. At
a minimum, they should include position verification, source uniformity, dwell
time function, indicator lamps and other status/operational displays, and
visual inspection for integrity of all applicators and catheters to be used for
the treatment.
i) Perform
tests following source or device exchange in accordance with the manufacturer's
instruction manual for:
1) Timer
accuracy/constancy, if appropriate;
2) Calibration
of the source output following the manufacturer's instructions; and
3) Interlock/interrupt
checks (i.e., interrupt test, cartridge lock test, emergency retraction test
and catheter connection test), if appropriate.
j) The
licensee shall retain a record of each item in subsections (b), (d), (e), (h)
and (i) for intravascular brachytherapy units for five years. The records shall
include:
1) The date of the
verification, inspection or check.
2) The
manufacturer's name, model and serial number of the intravascular brachytherapy
unit.
3) Results of the
verification, inspection or check.
4) Notations indicating
the operability of each component.
5) The identity of the
individual who performed the check.