32 Ill. Adm. Code 335.6010
Use of Sealed Sources for Diagnosis
Section 335
Section 335.6010 Use of
Sealed Sources for Diagnosis
a) A licensee shall use only sealed sources that are not in
medical devices for diagnostic medical uses if the sealed sources are approved
in the Sealed Source and Device Registry for diagnostic medicine. The sealed
sources may be used for diagnostic medical uses that are not explicitly listed
in the Sealed Source and Device Registry but shall be used in accordance with
the radiation safety conditions and limitations described in the Sealed Source
and Device Registry.
b)
A licensee shall only use medical
devices containing sealed sources for diagnostic medical uses if both the
sealed sources and medical devices are approved in the Sealed Source and Device
Registry for diagnostic medical uses. The diagnostic medical devices may be
used for diagnostic medical uses that are not explicitly listed in the Sealed
Source and Device Registry, but shall be used in accordance with the radiation
safety conditions and limitations described in the Sealed Source and Device
Registry.
c) Sealed
sources and devices for diagnostic medical uses may be used in research in
accordance
with an active Investigational Device Exemption (IDE)
application accepted by the U.S. Food and Drug Administration provided the
requirements of Section 335.35 are met.