32 Ill. Adm. Code 335.7090
Therapy-related Computer Systems for Manual Brachytherapy
Section 335.7090 Therapy-related Computer Systems for
Manual Brachytherapy
The licensee shall:
a) Perform
acceptance testing on the treatment planning system of therapy-related computer
systems in accordance with published protocols accepted by nationally
recognized bodies. At a minimum, the acceptance testing must include, as
applicable, verification of:
1) The
source-specific input parameters required by the dose calculation algorithm;
2) The
accuracy of dose, dwell time, and treatment time calculations at representative
points;
3) The
accuracy of isodose plots and graphic displays; and
4) The
accuracy of the software used to determine sealed source positions from
radiographic images.
b) Maintain
a record of acceptance testing and copies of the protocols used for acceptance
testing in accordance with this Section for 5 years after discontinuation of
use of the treatment planning system.