32 Ill. Adm. Code 335.8020
Installation, Maintenance, Adjustment and Repair Restrictions
Section 335
Section 335.8020 Installation,
Maintenance, Adjustment and Repair Restrictions
a) Only a person specifically licensed by the Agency, the U.S.
Nuclear Regulatory Commission, an Agreement State or a Licensing State shall
install, maintain, adjust or repair a remote afterloader unit, intravascular
brachytherapy unit, teletherapy unit or gamma stereotactic radiosurgery unit
that involves work on the sources shielding, the sources driving unit or other
electronic or mechanical component that could expose the sources, reduce the
shielding around the sources, or compromise the radiation safety of the unit or
the sources.
b) Except
for a low dose-rate remote afterloader unit and intravascular brachytherapy
unit, only a person specifically licensed by the Agency, the U.S. Nuclear
Regulatory Commission, an Agreement State or a Licensing State shall install,
replace, relocate, or remove a sealed source or source contained in other
remote afterloader units, intravascular brachytherapy units, teletherapy units
or gamma stereotactic radiosurgery units.
c) For
a low dose-rate remote afterloader unit and intravascular brachytherapy unit,
only a person specifically licensed by the Agency, the U.S. Nuclear Regulatory
Commission, an Agreement State or a Licensing State or an authorized medical
physicist shall install, replace, relocate or remove a sealed source contained
in the unit.
d) A
licensee shall retain a record of the installation, maintenance, adjustment and
repair of remote afterloader units, intravascular brachytherapy units,
teletherapy units and gamma stereotactic radiosurgery units as required by this
Section for 5 years. For each installation, maintenance, adjustment and
repair, the record must include the date, description of the service, names of
the individuals who performed the work, and a copy of the specific license
authorizing the service.