32 Ill. Adm. Code 335.8190
Full Calibration Measurements on Gamma Stereotactic Radiosurgery Units
Section 335.8190 Full Calibration Measurements on Gamma
Stereotactic Radiosurgery Units
a) A
licensee authorized to use a gamma stereotactic radiosurgery unit for medical
use shall perform full calibration measurements on each unit:
1) Before
the first medical use of the unit;
2) Before
medical use under the following conditions:
A) Whenever
spot-check measurements indicate that the output differs by more than 5 percent
from the output obtained at the last full calibration corrected mathematically
for radioactive decay;
B) Following
replacement of the sources or following reinstallation of the gamma
stereotactic radiosurgery unit in a new location; and
C) Following
any repair of the gamma stereotactic radiosurgery unit that includes removal of
the sources or major repair of the components associated with the source
assembly; and
3) At
intervals not exceeding 1 year, with the exception that relative helmet factors
need only be determined before the first medical use of a helmet and following
any damage to a helmet.
b) To
satisfy the requirement of subsection (a) of this Section, full calibration
measurements must include determination of:
1) The output within ± 3
percent;
2) Relative
helmet factors;
3) Isocenter
coincidence;
4) Timer accuracy
and linearity over the range of use;
5) On-off
error;
6) Trunnion
centricity;
7) Treatment
table retraction mechanism, using backup battery power or hydraulic backups
with the unit off;
8) Helmet
microswitches;
9) Emergency
timing circuits; and
10) Stereotactic
frames and localizing devices (trunnions).
c) A
licensee shall use the dosimetry system described in Section 335.8080(a) of
this Part to measure the output for one set of exposure conditions. The
remaining radiation measurements required in subsection (b)(1) of this Section
may be made using a dosimetry system that indicates relative dose rates.
d) A
licensee shall make full calibration measurements required by subsection (a) of
this Section in accordance with published protocols accepted by nationally
recognized bodies.
e) A
licensee shall mathematically correct the outputs determined in subsection
(b)(1) of this Section at intervals not exceeding 1 month for cobalt-60 and at
intervals consistent with 1 percent physical decay for all other radionuclides.
f) Full
calibration measurements required by subsection (a) of this Section and
physical decay corrections required by subsection (e) of this Section must be
performed by the authorized medical physicist.
g) A
licensee shall maintain a record of the gamma stereotactic radiosurgery unit
full calibrations required by this Section for 5 years.
h) The record must include:
1) The
date of the calibration;
2) The
manufacturer's name, model and serial number of the gamma stereotactic
radiosurgery units, the sources, and the instruments used to calibrate the
units;
3) The
results and an assessment of the full calibrations; and
4) The
signature of the authorized medical physicist who performed the full
calibration.