32 Ill. Adm. Code 335.8210
Additional Technical Requirements for Mobile Remote Afterloader Units
Section 335.8210 Additional Technical Requirements for Mobile Remote Afterloader Units
a) A licensee providing
mobile remote afterloader service shall:
1) Check
survey instruments before medical use at each address of use or on each day of
use, whichever is more frequent; and
2) Account
for all sources before departure from a client's address of use.
b) In
addition to the periodic spot-checks required by Section 335.8170 of this Part,
a licensee authorized to use mobile afterloaders for medical use shall perform
checks on each remote afterloader unit before use at each address of use. At a
minimum, checks must be made to verify the operation of:
1) Electrical
interlocks on treatment area access points;
2) Source
exposure indicator lights on the remote afterloader unit, on the control
console and in the facility;
3) Viewing
and intercom systems;
4) Applicators,
source transfer tubes and transfer tube-applicator interfaces;
5) Radiation
monitors used to indicate room exposures;
6) Source
positioning (accuracy); and
7) Radiation
monitors used to indicate whether the source has returned to a safe shielded
position.
c) In
addition to the requirements for checks in subsection (b) of this Section, a
licensee shall ensure overall proper operation of the remote afterloader unit
by conducting a simulated cycle of treatment before use at each address of use.
d) If
the results of the checks required in subsections (b) and (c) of this Section
indicate the malfunction of any system, a licensee shall lock the control
console in the off position and not use the unit except as may be necessary to
repair, replace or check the malfunctioning system.
e) A
licensee shall retain a record of each check for mobile remote afterloader
units required by this Section for 5 years.
f) The record must include:
1) The
date of the check;
2) The
manufacturer's name, model and serial number of the remote afterloader unit;
3) Notations
accounting for all sources before the licensee departs from a facility;
4) Notations
indicating the operability of each entrance door electrical interlock,
radiation monitors, source exposure indicator lights, viewing and intercom
system, applicators, source transfer tubes, transfer tube applicator
interfaces, and source positioning accuracy; and
5) The
signature of the individual who performed the check.