32 Ill. Adm. Code 335.8230
Therapy-related Computer Systems for Remote Afterloader Units, Teletherapy Units and Gamma Stereotactic Units
Section 335.8230 Therapy-related Computer Systems for
Remote Afterloader Units, Teletherapy Units and Gamma Stereotactic Units
The licensee shall perform acceptance testing on the
treatment planning system of therapy-related computer systems in accordance
with published protocols accepted by nationally recognized bodies. At a
minimum, the acceptance testing must include, as applicable, verification of:
a) The
source-specific input parameters required by the dose calculation algorithm;
b) The accuracy
of dose, dwell time and treatment time calculations at representative points;
c) The accuracy of isodose
plots and graphic displays;
d) The
accuracy of the software used to determine sealed source positions from
radiographic images; and
e) The accuracy
of electronic transfer of the treatment delivery parameters to the treatment
delivery unit from the treatment planning system.