32 Ill. Adm. Code 340.510
General
Section 340
Section 340.510 General
a) Each licensee or registrant shall make, or cause to be made,
surveys, including surveys of the subsurface, where appropriate:
1) That demonstrate compliance with this Part; and
2) That evaluate:
A) The extent of radiation levels;
B) Concentrations or quantities of radioactive material; and
C) The potential radiological hazards of radiation levels and
residual radioactivity detected.
b) The licensee or registrant shall ensure that instruments and
equipment used for quantitative radiation measurements (e.g., dose rate and
effluent monitoring) are calibrated at intervals not to exceed 12 months for
the radiation measured or at alternative intervals specified in regulations of
the Agency, the U.S. Nuclear Regulatory Commission, an Agreement State or a
Licensing State. To satisfy this requirement, the licensee shall:
1) Post a legible note on the instrument showing the date of
calibration; and
2) Ensure that instrument calibrations are performed by persons
specifically licensed by the Agency, the U.S. Nuclear Regulatory Commission, an
Agreement State or a Licensing State to perform such calibrations.
c) On each day of use, prior to using an instrument to perform
required monitoring, the licensee or registrant shall verify that the
instrument is operational. Operational checks for radiation measurement or
radiation detection instruments shall include verification of response to a
source of radiation.
d) Except for those dosimeters used to measure the dose to any
extremity, personnel dosimeters that require processing to determine the
radiation dose and that are used by licensees or registrants to comply with
Section 340.210, with other applicable provisions of 32 Ill. Adm. Code: Chapter
II, Subchapters b and d or with conditions specified in a license shall be
processed and evaluated by a qualified dosimetry processor. A dosimetry
processor is qualified if:
1) It holds current personnel dosimetry accreditation from the
National Voluntary Laboratory Accreditation Program (NVLAP) of the National
Institute of Standards and Technology; and
2) It is approved by NVLAP for the type of radiation or
radiations that most closely approximate the type of radiation or radiations
for which the individual wearing the dosimeter is monitored.
e) A licensee or registrant shall obtain Agency approval prior to
using pocket ionization chambers or electronic dosimeters to determine
radiation dose, to comply with Section 340.210, or with other applicable
provisions of 32 Ill. Adm. Code: Chapter II, Subchapters b and d or with
conditions specified in a license. The Agency will grant approval provided the
licensee or registrant submits information describing the type and range of the
dosimeters and describes a program to ensure the accuracy, reliability,
precision and security of the dosimetry data.
f) The licensee or registrant shall ensure that adequate
precautions are taken to prevent deceptive exposure of an individual monitoring
device.