32 Ill. Adm. Code 340.540
Calibration of Survey Instruments
Section 340.540 Calibration
of Survey Instruments
a) Unless specified in another Part, a licensee shall have each
survey instrument used to show compliance with this Part calibrated before
first use, annually and following a repair that affects the calibration. A
licensee shall:
1) Calibrate all scales with readings up to 10 mSv (1000 mrem)
per hour with a radiation source;
2) Calibrate two separated readings on each scale or decade that
will be used to show compliance; and
3) Conspicuously note on the instrument the date of calibration.
b) A licensee may not use survey instruments if the difference
between the indicated exposure rate and the calculated exposure rate is more
than 20 percent.