32 Ill. Adm. Code 350.1000
Requirements for Radiography Equipment Using Radiographic Exposure Devices
Section 350
Section 350.1000
Requirements for Radiography Equipment Using Radiographic Exposure Devices
a) Equipment used in industrial radiographic operations involving
the use of radiographic exposure devices shall meet the following minimum
criteria:
1) Each radiographic exposure device, source assembly, or sealed
source and all associated equipment:
A) Manufactured on or before July 1, 1994, and used after January
10, 1996, shall meet the requirements specified in American National Standards
Institute (ANSI) N432-1980, "Radiological Safety for the Design and
Construction of Apparatus for Gamma Radiography", published January 1981,
as NBS Handbook 136, exclusive of subsequent amendments or editions. This
publication may be purchased from the American National Standards Institute,
Inc., 25 West 43
rd
Street, New York NY 10036; Telephone: (212)
642-4900. However, equipment used in industrial radiographic operations need
not comply with section 8.9.2(c) of the Endurance Test in ANSI N432-1980, if
the prototype equipment has been tested using a torque value representative of
the torque that an individual using the radiography equipment can realistically
exert on the lever or crankshaft of the drive mechanism; and/or
B) Manufactured
after July 1, 1994, and used after January 10, 1996, shall meet the
requirements specified in ANSI N43.9-1991, "American National Standard for
Gamma Radiography – Specifications for Design and Testing of Apparatus",
published 1991, exclusive of subsequent amendments or editions.
2) Each radiographic exposure device shall have attached to it
one or more durable, legible, clearly visible labels bearing the:
A) Chemical symbol and mass number of the radionuclide in the
device;
B) Activity of the sealed source and the date this activity was
last measured;
C) Model and serial number of the sealed source;
D) Manufacturer of the sealed source; and
E) Licensee's name, address and telephone number.
3) Each radiographic exposure device intended for use as a Type B
transport container shall meet the applicable requirements of 32 Ill. Adm. Code
341.
4) Radiographic exposure devices, source assemblies, source
changers and associated equipment that allow the source to be moved out of the
device for routine operation shall meet the following additional requirements:
A) The coupling between the source assembly and the control cable
shall be designed in a manner that the source assembly will not become
disconnected if cranked outside the guide tube. The coupling shall be such
that it cannot be unintentionally disconnected under normal conditions.
B) The device shall automatically secure the source assembly when
it is cranked back into the shielded position within the device. This securing
system shall only be released by means of a deliberate operation of the
exposure device.
C) The outlet fittings, lock box and drive cable fittings on each
radiographic exposure device shall be equipped with safety plugs or covers,
which shall be installed during storage and transportation, to protect the
source assembly from water, mud, sand or other foreign matter.
D) Each sealed source or source assembly shall have attached to
it, or engraved in it, a durable, legible, visible label with the words:
"DANGER-RADIOACTIVE". The label shall not interfere with the safe
operation of the exposure device or associated equipment.
E) The guide tube, if manufactured on or before July 1, 1994, and
used after January 10, 1996, shall have passed a kinking test that closely
approximates the kinking forces likely to be encountered during use and the
crushing tests for the control units specified in ANSI N432-1980,
"Radiological Safety for the Design and Construction of Apparatus for
Gamma Radiography", published January 1981, as NBS Handbook 136, exclusive
of subsequent amendments or editions. Guide tubes manufactured after July 1,
1994, and used after January 10, 1996, shall have passed a kinking test that
closely approximates the kinking forces likely to be encountered during use and
the crushing tests for the control units specified in ANSI N43.9-1991,
"American National Standard for Gamma Radiography – Specifications for
Design and Testing of Apparatus", published 1991, exclusive of subsequent
amendments or editions.
F) Use of a guide tube shall be necessary to move the source out
of the device.
G) An exposure head, endcap or similar device designed to prevent
the source assembly from extending beyond the end of the guide tube shall be
attached to the outermost end of the guide tube during radiographic operations.
H) The guide tube exposure head connection, if these parts were
manufactured on or before July 1, 1994, and used after January 10, 1996, shall
be able to withstand the tensile test for control units specified in ANSI
N432-1980, "Radiological Safety for the Design and Construction of
Apparatus for Gamma Radiography", published January 1981, as NBS Handbook
136, exclusive of subsequent amendments or editions If these parts were
manufactured after July 1, 1994, and used after January 10, 1996, the guide
tube exposure head connection shall be able to withstand the tensile test for
control units specified in ANSI N43.9-1991, "American National Standard
for Gamma Radiography – Specifications for Design and Testing of
Apparatus", published 1991, exclusive of subsequent amendments or
editions.
I) Source changers shall provide a system for assuring that the
source will not be accidentally withdrawn from the changer when connecting or
disconnecting the drive cable to or from a source assembly.
b) Modification of any radiographic exposure device, source
assembly, source changer and associated equipment is prohibited unless the
Agency, the U.S. Nuclear Regulatory Commission or an Agreement State has
determined that the design of any replacement component, including source
holder, source assembly, control or guide tube would not compromise the design
safety features of the system.
c) Each radiographic exposure device, source changer and storage
container shall be provided with a lock or lockable outer container designed to
prevent unauthorized or accidental removal or exposure of a sealed source.
d) Each radiographic exposure device and each transport container
shall bear a permanent, durable, legible, clearly visible marking or label that
has, as a minimum, the standard radiation caution symbol, depicted in 32 Ill.
Adm. Code 340.Illustration A, and the following wording:
CAUTION (OR DANGER)
RADIOACTIVE MATERIAL
NOTIFY CIVIL AUTHORITIES (OR NAME OF
COMPANY)
In addition,
transport containers shall meet the applicable requirements of 32 Ill. Adm. Code
341.