32 Ill. Adm. Code 350.1010
Limits on Levels of Radiation for Radiographic Exposure Devices, Source Changers and Transport Containers
Section 350
Section 350.1010 Limits on
Levels of Radiation for Radiographic Exposure Devices, Source Changers and
Transport Containers
a) Radiographic exposure devices manufactured prior to July 1,
1994, shall not be used for industrial radiography unless they meet the
following minimum criteria:
1) Radiographic exposure devices and source changers measuring
less than 10 centimeters (4 inches) from the sealed source storage position to
any exterior surface of the device shall have no radiation level in excess of
12.9 µC/kg (50 mR) per hour at 15 centimeters (6 inches) from any exterior
surface of the device.
2) Radiographic exposure devices measuring a minimum of 10
centimeters (4 inches) from the sealed source storage position to any exterior
surface of the device shall not have radiation levels in excess of 2 mSv
(equivalent to 200 mrem, 200 mR, or 51.6 µC/kg) per hour at any exterior surface,
and 0.1 mSv (equivalent to 10 mrem, 10 mR, or 2.58 µC/kg) per hour at 1 meter
(39.4 inches) from any exterior surface.
3) The radiation levels specified in subsections (1) and (2) of
this Section shall be determined with the sealed source in the shielded
position (i.e., "off").
b) Radiographic exposure devices, source changers and transport
containers manufactured on or after July 1, 1994, and used after January 10,
1996, shall not have a maximum exposure rate in excess of 2 mSv (200 mrem) per
hour at any exterior surface and 0.1 mSv (10 mrem) per hour at 1 meter from any
exterior surface with the sealed source in the shielded position.