32 Ill. Adm. Code 350.1040
Radiation Survey Instruments
Section 350
Section 350.1040 Radiation
Survey Instruments
a) The licensee or registrant shall maintain sufficient
calibrated and operable radiation survey instruments to make physical radiation
surveys as required by this Part and 32 Ill. Adm. Code 340.510(a). Instrumentation
required by this Section shall have a range that 0.02 mSv (0.516 µC/kg, 2 mrem
or 2 mR) per hour through 0.01 Sv (258 µC/kg or 1 R) per hour can be measured.
b) Each radiation survey instrument shall be calibrated:
1) At energies appropriate for use;
2) At intervals not to exceed 6 months and after each instrument
servicing other than battery replacement;
3) To accuracy that can be demonstrated within plus or minus 20
percent;
4) At two or more widely separated points, other than zero, on
each scale, or one point of each scale for digital devices. For instruments
without multiple scales, calibration shall be performed at six points equally
spaced across the range of 0.02 mSv (0.516 µC/kg 2 mrem or 2 mR) per hour to 0.01
Sv (258 µC/kg, 1 rem or 1 R) per hour; and
5) By a person licensed by the Agency, the U.S. Nuclear
Regulatory Commission, an Agreement State or a Licensing State to perform such
service.
c) Records of calibrations shall be maintained for 5 years after
the calibration date for inspection by the Agency.
d) Immediately prior to use, a radiation survey instrument shall
be checked to ensure that it is operating properly by bringing it near a source
of radiation and observing a response. Instruments that fail to respond shall
not be used.