32 Ill. Adm. Code 350.1090
Permanent Radiographic Installations
Section 350
Section 350.1090 Permanent
Radiographic Installations
Permanent radiographic
installations using radiographic exposure device(s) having high radiation area
entrance controls of the type described in 32 Ill. Adm. Code 340.610(a)(2),
340.610(a)(3) and 340.610(b) shall also meet the following requirements:
a) Each entrance that is used for personnel access to the high
radiation area shall have both visible and audible warning signals to warn of
the presence of radiation. The visible signal shall be activated by radiation.
The audible signal shall be activated when an attempt is made to enter the
installation while the source is exposed.
b) The entrance control device or alarm system shall be tested
for proper operation prior to beginning operations on each day of use. The
radiography system shall not be used if any entrance control device or alarm
system is operating improperly. If an entrance control device or alarm system
is operating improperly, it shall be labeled as defective immediately and
repaired. Before the radiography system is returned to service, the radiation
safety officer shall retest the entrance control device or alarm system and
approve the repair.
c) Records of tests performed pursuant to subsection (b) of this
Section shall be maintained for inspection by the Agency for 5 years.