32 Ill. Adm. Code 351.1050
Testing for Leakage or Contamination of Sealed Sources
Section 351
Section 351.1050 Testing for
Leakage or Contamination of Sealed Sources
a) Testing for leakage or contamination of sealed sources shall
be performed in accordance with 32 Ill. Adm. Code 340.410. Test samples shall
be taken from the surfaces of sources or source holders or from the surfaces of
devices in which sources are mounted and on which one might expect
contamination to accumulate. The wipe of a sealed source must be performed
using a leak test kit approved by the U.S. Nuclear Regulatory Commission, an
Agreement State or a Licensing State.
b) Each
energy compensation source that is not exempt from testing for leakage or
contamination in accordance with 32 Ill. Adm. Code 340.410(b) shall be tested
at intervals not to exceed 3 years. In the absence of a certificate from a
transferor that a test has been made within the 3 years before the transfer,
the energy compensation source shall not be used until tested in accordance
with 32 Ill. Adm. Code 340.410.
c) If
a sealed source is found to be leaking in accordance with 32 Ill. Adm. Code
340.410, the licensee shall check the equipment associated with the leaking
source for radioactive contamination and, if contaminated, have it
decontaminated or disposed of by persons specifically authorized by the Agency,
the U.S. Nuclear Regulatory Commission, an Agreement State or a Licensing State
to perform such services.
d) The
licensee shall maintain records of testing for leakage or contamination of
sealed sources in accordance with 32 Ill. Adm. Code 340.1135.