32 Ill. Adm. Code 360.130
Electronic Brachytheraphy
Section 360.130 Electronic Brachytheraphy
a) Applicability.
Electronic brachytherapy devices shall be subject to the requirements of this
Section and shall be exempt from the requirements of Section 360.110, unless
otherwise noted in this Section.
1) An
electronic brachytherapy device that does not meet the requirements of this
Section shall not be used for irradiation of patients; and
2) An
electronic brachytherapy device shall only be utilized for human use
applications specifically approved by the U.S. Food and Drug Administration
(FDA) unless participating in a research study approved by the registrant's
Institutional Review Board.
b) Possession
of Survey Instruments. Each registrant using an electronic brachytherapy
device in accordance with this Section shall possess appropriately calibrated
portable monitoring equipment. At a minimum, the monitoring equipment shall
include a portable radiation measurement survey instrument capable of measuring
dose rates over the range 10 µSv (1 mrem) per hour to 10 mSv (1000 mrem) per
hour. The survey instruments shall be operable and calibrated within the prior
12 months for the applicable electronic brachytherapy source energy.
c) Facility
Design Requirements for Electronic Brachytherapy Devices. Each electronic
brachytherapy installation shall be provided with such primary and secondary
barriers as are necessary to assure compliance with 32 Ill. Adm. Code 340.
1) If
applicable, provision shall be made to prevent simultaneous operation of more
than one therapeutic radiation machine in a treatment room.
2) Access
to the treatment room shall be controlled by a door at each entrance.
3) Each
treatment room shall have provisions to permit continuous aural communication
and visual observation of the patient from the treatment control panel during
irradiation. The electronic brachytherapy device shall not be used for patient
irradiation unless the patient can be observed.
d) Control
Panel Functions. The control panel, in addition to the displays required by
other provisions in this Section, shall:
1) Provide
an indication of whether electrical power is available at the control panel and
if activation of the electronic brachytherapy source is possible;
2) Provide an indication
of whether x‑rays are being produced;
3) Provide
a means for indicating electronic brachytherapy source potential and current;
4) Provide a means for
terminating an exposure at any time; and
5) Include
an access control (locking) device that will prevent unauthorized use of the
electronic brachytherapy device.
e) Timer.
A suitable irradiation control device (timer) shall be provided to terminate
the irradiation after a pre-set time interval or integrated charge on a
dosimeter-based monitor. The timer shall operate according to the manufacturer's
design specifications.
f) Therapeutic
Radiological Physicist Support. The services of a therapeutic radiological
physicist shall be required in facilities having electronic brachytherapy
devices. The therapeutic radiological physicist shall be responsible for:
1) Evaluation of the
output from the electronic brachytherapy source;
2) Generation of the
necessary dosimetric information;
3) Supervision
and review of treatment calculations prior to initial treatment of any
treatment site;
4) Establishing
the periodic and day-of-use quality assurance checks and reviewing the data
from those checks as required in subsection (j);
5) Consultation with the
physician in treatment planning, as needed;
6) Performing
calculations/assessments regarding patient treatments that may constitute a
misadministration.
7) Determination
of the need for shielding or safe distances for individuals in the room during
electronic brachytherapy treatments, in accordance with the radiation dose
limits of 32 Ill. Adm. Code Part 340;
8) Implementation
of the use of shielding or safe distances as determined in subsection (f)(7).
g) Operating Procedures
1) Only
individuals approved by the physician or therapeutic radiological physicist
shall be present in the treatment room during treatment.
2) Electronic
brachytherapy devices shall not be made available for medical use unless the
requirements of this Section have been met.
3) The
electronic brachytherapy device shall be inoperable, either by hardware or
password, when unattended by qualified staff or service personnel.
4) During
operation, the therapeutic radiologic physicist shall ensure that all persons
in the treatment room, and all persons entering the treatment room, are
prevented from exceeding the radiation dose limits of 32 Ill. Adm. Code Part 340.
5) If a
patient must be held in position during treatment, mechanical supporting or
restraining devices shall be used.
6) Written
procedures shall be developed, implemented, and maintained for responding to an
abnormal situation. These procedures shall include:
A) Instructions
for responding to equipment failures and the names of the individuals
responsible for implementing corrective actions; and
B) The
names and telephone numbers of the physician and the therapeutic radiological physicist
to be contacted if the device or console operates abnormally.
7) A
copy of the current operating and emergency procedures shall be physically
located at the electronic brachytherapy device control console.
8) Instructions
shall be posted at the electronic brachytherapy device control console to
inform the electronic brachytherapy device operator of the names and telephone
numbers of the physician and the therapeutic radiological physicist to be
contacted if the device or console operates abnormally.
h) Safety Precautions for
Electronic Brachytherapy Devices
1) A
therapeutic radiological physicist shall determine which persons in the
treatment room require radiation monitoring when the beam is energized.
2) A
physician and a therapeutic radiological physicist shall be physically present
during the initiation of all patient treatments involving the electronic
brachytherapy device.
3) A
therapeutic radiological physicist and either a physician or an electronic
brachytherapy device operator, under the supervision of a physician, who has
been trained in the operation of, and emergency response for, the electronic
brachytherapy device, shall be physically present during continuation of all
patient treatments involving the electronic brachytherapy device.
4) A
therapeutic radiological physicist shall designate shield locations or safe
distances sufficient to meet the requirements of 32 Ill. Adm. Code 340 for any
individual, other than the patient, in the treatment room;
5) All
personnel in the treatment room are required to remain behind shielding or at a
safe distance specified by the therapeutic radiological physicist during
treatment. A therapeutic radiological physicist shall approve any deviation
from this requirement and shall designate alternative radiation safety
protocols, compatible with patient safety, to provide an equivalent degree of
protection.
i) Electronic
Brachytherapy Source Calibration Measurements
1) Calibration
of the electronic brachytherapy source output for an electronic brachytherapy
device shall be performed by, or under the direct supervision of, a therapeutic
radiological physicist.
2) Calibration
of the electronic brachytherapy source output shall be made for each electronic
brachytherapy source, or after any repair affecting the x-ray beam generation,
or when indicated by the electronic brachytherapy source quality assurance
checks.
3) Calibration
of the electronic brachytherapy source output shall utilize a dosimetry system
that meets the requirements of subsection 360.110(d)(4).
4) Calibration
of the electronic brachytherapy source output shall include, as applicable,
determination of:
A) The
output within 2% of the expected value, if applicable, or determination of the
output if there is no expected value;
B) Timer accuracy and
linearity over the typical range of use;
C) Proper operation of
back-up exposure control devices;
D) Evaluation
that the relative dose distribution about the source is within 5% of that
expected; and
E) Source
positioning accuracy to within one millimeter within the applicator.
5) Calibration
of the x-ray source output shall be in accordance with the manufacturer's
calibration protocol.
6) The
registrant shall maintain a record of each calibration in an auditable form for
5 years. The record shall include: the date of the calibration; the
manufacturer's name, model number and serial number for the electronic
brachytherapy device and a unique identifier for its electronic brachytherapy
source; the model numbers and serial numbers of the instruments used to
calibrate the electronic brachytherapy device; and the name and signature of
the therapeutic radiological physicist responsible for performing the
calibration.
j) Periodic
and Day-of-Use Quality Assurance Checks for Electronic Brachytherapy Devices
1) Quality
assurance checks shall be performed on each electronic brachytherapy device:
A) At the beginning of each
day of use;
B) Each time the device is
moved to a new room or site; and
C) After each x-ray tube
installation.
2) The
registrant shall perform periodic quality assurance checks required by
subsection (j)(1) in accordance with procedures established by the therapeutic
radiological physicist;
3) Quality
assurance checks shall include, at a minimum:
A) Verification
that output of the electronic brachytherapy source falls within 3% of expected
values, as appropriate for the device, as determined by:
i) Output as a function
of time; or
ii) Output as a function
of setting on a monitor chamber.
B) Verification
of the consistency of the dose distribution to within 3% of that found during
calibration required by subsection (i); and
C) Validation
of the operation of positioning methods to ensure that the treatment dose
exposes the intended location within 1 mm.
4) The
registrant shall use a dosimetry system that has been intercompared within the
previous 12 months with the dosimetry system described in subsection (i)(3) to
make the quality assurance checks required in this subsection (j).
5) The
registrant shall review the results of each radiation output quality assurance
check according to the following procedures:
A) A
physician and therapeutic radiological physicist shall be immediately notified
if any parameter is not within its acceptable tolerance. The electronic
brachytherapy device shall not be made available for subsequent medical use
until the therapeutic radiological physicist has determined that all parameters
are within their acceptable tolerances; and
B) The
therapeutic radiological physicist shall review and sign the results of each
radiation output quality assurance check at intervals not to exceed 30 days.
6) Quality
assurance checks shall, at a minimum, assure:
A) Proper
operation of radiation exposure indicator lights on the electronic
brachytherapy device and on the control console;
B) Proper
operation of viewing and intercom systems in each electronic brachytherapy
facility, if applicable;
C) Proper operation of
radiation monitors, if applicable;
D) The
integrity of all cables, catheters or parts of the device that carry high
voltages; and
E) Connecting
guide tubes, transfer tubes, transfer-tube-applicator interfaces, and treatment
spacers are free from any defects that interfere with proper operation.
7) If
the results of the safety device quality assurance checks indicate the
malfunction of any system, a registrant shall secure the control console in the
OFF position and not use the electronic brachytherapy device except as may be
necessary to repair, replace, or check the malfunctioning system.
8) The
registrant shall maintain a record of each quality assurance check in an
auditable form for 3 years.
A) The
record shall include the date of the quality assurance check; the
manufacturer's name, model number and serial number for the electronic
brachytherapy device; the name and signature of the individual who performed
the periodic quality assurance check; and the name and signature of the
therapeutic radiological physicist who reviewed the quality assurance check;
and
B) The
record shall also include the unique identifier for the electronic
brachytherapy source; the manufacturer's name; and the model number and serial
number for the instruments used to measure the radiation output of the electronic
brachytherapy device.
k) Therapy-Related
Computer Systems. The registrant shall perform acceptance testing on the
treatment planning system of electronic brachytherapy-related computer systems
in accordance with the manufacturer's acceptance testing protocol.
1) Acceptance
testing shall be performed by, or under the direct supervision of, a
therapeutic radiological physicist. At a minimum, the acceptance testing shall
include, as applicable, verification of:
A) The
source-specific input parameters required by the dose calculation algorithm;
B) The
accuracy of dose, dwell time, and treatment time calculations at representative
points;
C) The accuracy of isodose
plots and graphic displays;
D) The
accuracy of the software used to determine radiation source positions from
radiographic images; and
E) If the
treatment planning system is different from the treatment delivery system, the
accuracy of electronic transfer of the treatment delivery parameters to the
treatment delivery unit from the treatment planning system.
2) The
position indicators in the applicator shall be compared to the actual position
of the source or planned dwell positions, as appropriate, at the time of
commissioning.
3) Prior
to each patient treatment regimen, the parameters for the treatment shall be
evaluated and approved by the physician and the therapeutic radiological
physicist for correctness through means independent of that used for the
determination of the parameters.
l) Training
1) A
registrant shall provide instruction, initially and at least annually, to all
individuals who operate the electronic brachytherapy device, as appropriate to
the individual's assigned duties, in the operating procedures identified in
subsection (g). If the interval between patients exceeds one year, retraining
of the individuals shall be provided.
2) Physicians,
therapeutic radiological physicists, and electronic brachytherapy device
operators shall receive device specific instruction initially from the
manufacturer and annually from either the manufacturer or other qualified
trainer. The training shall be of a duration recommended by the manufacturer's
training protocol. The training shall include, but not be limited to:
A) Device-specific
radiation safety requirements;
B) Device operation;
C) Clinical use for the
types of use approved by the FDA;
D) Emergency procedures,
including an emergency drill; and
E) The registrant's quality
assurance program.
3) A
registrant shall retain a record of individuals receiving instruction required
by this subsection (l) for 3 years. The record shall include a list of the
topics covered, the date of the instruction, the names of the attendees, and
the names of the individuals who provided the instruction.
m) Mobile
Electronic Brachytherapy Service. A registrant providing mobile electronic
brachytherapy service shall, as a minimum:
1) Check
all radiation survey instruments before medical use at each address of use or
on each day of use, whichever is more restrictive.
2) Account
for the electronic brachytherapy source in the electronic brachytherapy device
before departure from the client's address.
3) Perform,
at each location on each day of use, all of the required quality assurance
checks specified in subsection (j) to assure proper operation of the device.