32 Ill. Adm. Code 360.20
Definitions
Section 360
Section 360.20 Definitions
As used in this Part, the
following definitions apply:
"Accelerator" (also "particle
accelerator") means any therapeutic machine capable of producing a useful
beam of x-rays or charged particles with energies of 1 MeV or greater.
Accelerators include cyclotrons, betatrons and linear accelerators.
"Accelerator facility" means the location at which
one or more particle accelerators are installed and are operated under the same
administrative control.
"Agency" means the Illinois Emergency Management
Agency.
"Aluminum equivalent" means the thickness of type
1100 aluminum alloy affording the same attenuation, under specified conditions,
as the material in question. The nominal chemical composition of type 1100
aluminum alloy is 99.00 percent minimum aluminum, 0.12 percent copper.
"Applicator" means a structure which determines the
extent of the treatment field at a given distance from the source of the beam.
"Attenuation block" means a block or stack, having
dimensions 20 centimeters by 20 centimeters by 3.8 centimeters, of aluminum
equivalent. Copper may be substituted for aluminum if an appropriate thickness
is used for the kVp selected, as indicated below:
kVp
Millimeters
of Copper Equivalent to 3.8 centimeters of aluminum
99 or less
2.0
100 to 125
2.5
greater than 125
3.0
"Automatic exposure control" means a device that
automatically controls one or more technique factors in order to obtain at a
preselected location or locations a required quantity of radiation (see
"Phototimer").
"Barrier" (see "Protective barrier").
"Beam" means a flow of electromagnetic or
particulate radiation that passes through the opening in the beam limiting
device and that is used for diagnosis or treatment.
"Beam axis" (see "Central axis of the
beam").
"Beam-limiting device" means a device that provides
a means to restrict the dimensions of the x-ray field (see
"Collimator", "Diaphragm" and "Shutter").
"Beam monitoring system" means a system of devices
that will monitor the useful beam during irradiation and will terminate
irradiation when a preselected number of monitor units has been accumulated.
"Beam scattering filter" means a filter placed in
an electron beam in order to scatter the beam and provide a more uniform
distribution of electrons in the beam.
"Central axis of the beam" means the line passing
through the source of the beam and the center of the plane formed by the edge
of the first beam-limiting device.
"Charged particle beam" (see "Beam").
"Coefficient of variation" means the ratio of the
standard deviation to the mean value of a population of observations.
"Collimator" means a device or mechanism by which
the x-ray beam is restricted in size (see "Beam-limiting device").
"Computed tomography" or "CT" means the
production of a tomogram by the acquisition and computer processing of x-ray
transmission data.
"Computed tomography dose index" or "CTDI"
means the integral of the dose profile along a line perpendicular to the
tomographic plane divided by the product of the nominal tomographic section
thickness and the number of tomograms produced in a single scan.
"Contact
hour" means the number of hours an individual is in contact with an
instructor. One contact hour equals 50 minutes.
"Contact therapy system" means an x-ray system used
for therapy that is designed for very short treatment distances (5 centimeters
or less), usually employing peak tube potentials in the range of 20 to 50 kVp.
"Control panel" means that part or parts of the
x-ray system upon which are mounted the switches, knobs, pushbuttons and other
hardware necessary for setting the technique factors prior to initiating an
x-ray exposure.
"CT gantry" means the tube housing assemblies,
beam-limiting devices, detectors and the supporting structures and frames that
hold these components.
"Dead-man switch" means a switch constructed so
that a circuit-closing contact can be maintained only by continuous pressure on
the switch by the operator.
"Diagnostic imaging specialist" means a person who
possesses the knowledge, training and experience to apply the principles of
radiological physics to diagnostic x-ray applications. The diagnostic imaging
specialist shall be approved and registered by the Agency pursuant to 32 Ill.
Adm. Code 410.
"Diagnostic source assembly" means an x-ray tube
housing assembly, designed for use in diagnostic x-ray applications, with a
beam-limiting device attached.
"Diaphragm" means a device or mechanism by which
the x-ray beam is restricted in size (see "Beam-limiting device").
"Direct supervision" means an individual is in the
physical presence of a licensed practitioner who assists, evaluates and
approves of the individual's performance of the various tasks involved in the
application of ionizing radiation.
"Electronic
brachytherapy" means a method of radiation therapy in which an
electrically generated source of ionizing radiation is placed in or near the
tumor or target tissue to deliver a therapeutic radiation dosage.
"Electronic
brachytherapy device" means the system used to produce and deliver
therapeutic radiation, including x-ray tube, the control mechanism, the cooling
system and the power source.
"Electronic
brachytherapy device operator" means a radiation therapist accredited in
accordance with 32 Ill. Adm. Code 401 or a physician.
"Electronic
brachytherapy source" means the x-ray tube component used in an electronic
brachytherapy device.
"Field flattening filter" means a filter used to
provide dose uniformity over the area of a useful beam of x-rays at a specified
depth.
"Filter" means material placed in the useful beam
to absorb, preferentially, radiations based on energy level or to modify the
spatial distribution of the beam.
"Gantry" means that part of the system supporting
and allowing possible movements of the radiation head.
"General purpose x-ray system" means any
radiographic x-ray system that, by design, is not limited to radiographic
examination of specific anatomical regions.
"Gonad shield" means a protective device for the
testes or ovaries that provides a minimum of 0.5 millimeter lead equivalent
protection.
"Half-value layer" or "HVL" means the
thickness of a specified material that attenuates the beam of radiation to an
extent such that the exposure rate is reduced to one-half of its original
value.
AGENCY NOTE:
The contribution of all scattered radiation, other than any that might be
present initially in the beam concerned, should be minimized.
"Healing arts screening" means the examination of
human beings using x-ray machines for the detection or evaluation of potential
diseases when the examinations are not specifically ordered by a licensed
practitioner of the healing arts legally authorized to prescribe such x-ray
examinations for the purpose of diagnosis or treatment. However, healing arts
screening does not include mammography on self-referred patients.
"Image intensifier" means a device, installed in a
housing, that converts an x-ray pattern into a corresponding light image,
usually by electronic means.
"Image receptor" means any device, such as a
fluorescent screen or radiographic film, that transforms incident x-ray photons
either into a visible image or into another form that can be made into a
visible image by further transformations.
"Institutional
review board" means a committee that has been formally designated by the
registrant to approve, monitor and review biomedical and behavioral research
involving humans.
"Interlock" means a device arranged or connected
such that the occurrence of an event or condition is required before a second
event or condition can occur or continue to occur.
"Isocenter" means a fixed point in space located at
the center of the smallest sphere through which the central axis of the useful
beam passes at any beam orientation.
"Kilovolts peak" or "kVp" means the crest
value, in kilovolts, of the electric potential applied to the x-ray tube
between the cathode and anode of a pulsating electric potential generator.
"Lead equivalent" means the thickness of lead
affording the same attenuation, under specified conditions, as the material in
question.
"Leakage radiation" means all radiation emanating
from the diagnostic source assembly except for:
The useful
beam; and
The radiation
produced when the exposure switch or timer is not activated.
"Leakage technique factors" means the technique
factors used to measure leakage radiation from the diagnostic source assembly.
They are defined as follows:
For capacitor
energy storage equipment, the maximum-rated peak tube potential and the
maximum-rated number of exposures in 1 hour for operation at the maximum-rated
peak tube potential with the quantity of charge per exposure being 10
millicoulombs, i.e., 10 milliampere-seconds, or the minimum obtainable from the
unit, whichever is larger.
For field
emission equipment rated for pulsed operation, the maximum-rated peak tube
potential and the maximum-rated number of x-ray pulses in 1 hour for operation
at the maximum-rated peak tube potential.
For all other
equipment, the maximum-rated peak tube potential and the maximum-rated
continuous tube current for the maximum-rated peak tube potential.
"Light
field" means that area of the intersection of the light beam from the
beam-limiting device and any one of the sets of planes parallel to and including
the plane of the image receptor. The edge of the light field is defined as the
locus of points at which the illumination is 25 percent of that at the center
of the light field.
"Medical
event" means an event that meets the criteria in Section 360.120(i)(3).
"Medical radiographer" means a person other than a
licensed practitioner, accredited in accordance with the provisions of 32 Ill.
Adm. Code 401, or an individual exempt from the provisions of 32 Ill. Adm. Code
401, who performs medical radiation procedures and applies x-radiation, to any
part of the human body, for diagnostic purposes while under the supervision of
a licensed practitioner.
"Mobile equipment" (see "X-ray
equipment").
"Monitor unit" means a unit response from the beam
monitoring system from which the absorbed dose can be calculated.
"Moving beam therapy" means radiation therapy in
which there is displacement of the useful beam relative to the patient. Moving
beam therapy includes arc therapy, skip therapy and rotational beam therapy.
"Operator" means an individual who applies ionizing
radiation for diagnostic or therapeutic purposes.
"Phototimer" means a method for controlling
radiation exposures to image receptors by the amount of radiation that reaches
a radiation monitoring devices. The radiation monitoring devices is part of an
electronic circuit that controls the duration of time the tube is activated
(see "Automatic exposure control").
"Portable equipment" (see "X-ray
equipment").
"Portable
x-ray service provider" means a registrant who, under a physician's
authorization, provides x-ray procedures with hand-held or mobile radiographic
equipment in a patient's place of residence.
"Position indicating device" means a device on
intraoral dental x-ray equipment used to indicate the beam position and to
establish a definite source-skin distance.
"Prescribed dose" means
the total dose and dose per fraction as documented in the written directive.
The prescribed dose is an estimation from measured data from a specified
therapeutic machine using assumptions that are clinically acceptable for that
treatment technique and historically consistent with the clinical calculations
previously used for patients treated with the same clinical technique.
"Primary protective barrier" (see "Protective
barrier").
"Protective apron" means an apron of radiation
absorbing materials, at least 0.25 millimeter lead equivalent, used to reduce
exposure from leakage and scatter radiation.
"Protective barrier" means a barrier of radiation
absorbing materials used to reduce radiation dose. The types of protective
barriers are as follows:
"Primary
protective barrier" means the material, excluding filters, placed in the
useful beam to reduce the radiation dose.
"Secondary
protective barrier" means a barrier sufficient to attenuate the leakage
and scatter radiation to the required degree.
"Protective glove" means a glove made of radiation
absorbing materials, at least 0.25 millimeter lead equivalent, used to reduce
dose from leakage and scatter radiation.
"Radiation beam" (see "Beam").
"Radiation therapy simulation system" means a
radiographic/ fluoroscopic x-ray system used exclusively for localizing the
volume to be exposed during radiation therapy and confirming the position and
size of the therapeutic irradiation field.
"Radiologist
assistant" means a person, other than a licensed practitioner, who, as a
medical radiographer with advanced-level training and certification, performs a
variety of activities under the supervision of a radiologist certified by the
American Board of Radiology or the American Osteopathic Board of Radiology, in
the areas of patient care, patient management, clinical imaging and
interventional procedures. The radiologist assistant may not interpret images,
make diagnoses or prescribe medications or therapies.
"Reference plane" means a plane that is displaced
from and parallel to the tomographic plane.
"Scan" means the complete process of collecting
x-ray transmission data for the production of a tomogram. Data can be collected
simultaneously during a single scan for the production of one or more
tomograms.
"Scan increment" means the amount of relative
displacement of the patient support device with respect to the CT x-ray system
between successive scans measured along the direction of such displacement.
"Scatter radiation" means radiation that, during
passage through matter, has been deviated in direction.
"Secondary protective barrier" (see
"Protective barrier").
"Shadow tray" means a device attached to the radiation
head to support auxiliary beam-limiting material.
"Shutter" means an adjustable beam-limiting or
attenuating device, usually made of lead, fixed to an x-ray tube housing to
intercept or collimate the useful beam (see "Beam-limiting device").
"SID" means source-image receptor distance (see
"Source-image receptor distance").
"Source" means the focal spot of the x-ray tube.
"Source-image receptor distance" means the distance
from the source to the center of the input surface of the image receptor.
"Source to skin distance" or "SSD" means
the distance measured along the central ray from the center of the front
surface of the x-ray focal spot to the surface of the irradiated object.
"Special purpose x-ray system" means any
radiographic x-ray system that, by design, is limited to radiographic
examination of a specific anatomical region or to the extremities collectively.
"Spot film" means a radiograph that is made during
a fluoroscopic examination to permanently record conditions that exist during
that fluoroscopic procedure.
"Stationary beam therapy" means radiation therapy
in which there is no displacement of the useful beam relative to the patient
during irradiation.
"Stationary equipment" (see "X-ray
equipment").
"Supervision" means responsibility for and control
of quality, radiation safety and protection, and technical aspects of the
application of ionizing radiation to human beings for diagnostic and/or
therapeutic purposes.
"Technique factors" means the electrical potential
(kilovolts), current (milliamperes), exposure time parameters (seconds or
pulses) or a combination thereof, selectable at the control panel of an x-ray
system (see "Control panel").
"Therapeutic radiological physicist" means an
individual who has the knowledge, training and experience to measure ionizing
radiation, evaluate safety techniques, advise regarding radiation protection needs
and apply the principles of radiological physics to clinical radiation
therapy. The therapeutic radiological physicist shall be approved and
registered by the Agency pursuant to 32 Ill. Adm. Code 410.
"Tomogram" means the depiction of the x-ray attenuation
properties of a section through the body.
"Tomographic plane" means that geometric plane that
is identified as corresponding to the output tomogram.
"Tomographic section" means the volume of an object
whose x-ray attenuation properties are imaged in a tomogram.
"Useful beam" (see "Beam").
"X-ray equipment" means an x-ray system, sub-system
or component thereof. Types of x-ray equipment are as follows:
"Mobile
x-ray equipment" means x-ray equipment mounted on a permanent base with
wheels and/or casters for moving while completely assembled. Mobile x-ray
equipment includes x-ray equipment permanently mounted in vehicles.
"Portable
x-ray equipment" means x-ray equipment designed to be hand-carried.
"Stationary x-ray equipment" means x-ray equipment that
is installed in a fixed location.
"X-ray field" means, for diagnostic purposes, that
area of the intersection of the useful beam and any one of the set of planes
parallel to and including the plane of the image receptor. The edge of the
x-ray field is defined as the locus of points at which the exposure is 25
percent of that at the center of the x-ray field.
"X-ray system" means an assemblage of components
for the controlled production of x-rays. It includes minimally an x-ray
high-voltage generator, an x-ray control panel, an x-ray tube housing assembly,
a beam-limiting device and the necessary supporting structures. Additional
components that function with the system are considered integral parts of the
system. X-ray systems include diagnostic systems, therapeutic systems and accelerator
systems.