32 Ill. Adm. Code 360.TABLE B

B Half-Value Layer as a Function of Tube Potential

Last amended: 1998Year: 2026Length: 202 wordsOfficial source
Section 360 Section 360.TABLE B   Half-Value Layer as a Function of Tube Potential X-ray Tube Voltage (kilovolt peak) Minimum HVL (mm of Al) 1 Designed operating range Measured Operating Potential Specified Dental Systems 2 Other X-Ray Systems 3 Below 50 30 1.5 0.3 40 1.5 0.4 49 1.5 0.5 50 to 70 50 1.5 1.2 60 1.5 1.3 70 1.5 1.5 Above 71 71 2.1 2.1 80 2.3 2.3 90 2.5 2.5 100 2.7 2.7 110 3.0 3.0 120 3.2 3.2 130 3.5 3.5 140 3.8 3.8 150 4.1 4.1 1 Linear extrapolation or interpolation may be made for an x-ray tube potential (kVp) not listed in the table above (e.g., in the column entitled "Other X-ray Systems" operated at 20 kVp and 95 kVp, the minimum HVL required would be 0.2 and 2.6 millimeters of aluminum respectively). 2 "Specified Dental Systems" means any dental x-ray system designed for use with intraoral image receptors and manufactured after December 1, 1980. 3 "Other X-Ray Systems" means all x-ray systems required to meet the provisions of Sections 360.50, 360.60, 360.75, 360.90 (except "Specified Dental Systems") and 360.100 of this Part.  Half-value layer requirements for mammography systems are specified in Section 360.71(e) of this Part.