32 Ill. Adm. Code 380.60
Operating Procedures and Instructions
Section 380
Section 380.60 Operating
Procedures and Instructions
a) Individuals having access to analytical x-ray machines or
equipment shall be provided with specific written instructions concerning the
radiation hazards, safe working practices, and made aware of the symptoms of an
acute localized exposure to radiation. These instructions shall be posted near
the controls of the x-ray machine.
b) Medical personnel examining work-connected injuries shall be
informed of the possibility of radiation exposure to the worker from the
devices regulated in this Part.
c) Operators shall be instructed in the procedures for reporting
an actual or suspected radiation overexposure. When it has been determined
that an overexposure to an individual has occurred, it shall be reported to the
Agency without undue delay.
d) In cases where the primary x-ray beam is not intercepted by
the experimental apparatus under all conditions of operation, protective
measures shall be provided, such as auxiliary shielding, to avoid exposure to
the primary x-ray beam.
e) If, for any reason, it is necessary to temporarily
intentionally alter safety devices, such as bypassing interlocks or removing
shielding, such action shall be:
1) specified in writing and posted near the x-ray tube housing so
that other persons will know the existing status of the machine; and
2) terminated as soon as possible.
f) Whenever possible, an interlocking device that prevents the
entry of any portion of an individual body or extremities into the primary
beam, or causes the primary beam to be shut off upon entry into its path, shall
be provided.
g) Unused tube ports shall be closed in such a fashion that
accidental opening is not possible.