32 Ill. Adm. Code 422.140
Device Protocol
Section 422
Section 422.140 Device
Protocol
a) Quality Assurance
1) Professional
licensees providing measurement services using radon and radon product measurement
devices shall establish and maintain a Quality Assurance Program (QAP). These
programs shall include written procedures for attaining quality assurance
objectives and a system for recording and monitoring the results of the quality
assurance measurements for each device used. The QAP shall include the
maintenance of control charts and related statistical data.
2) The
objective of quality assurance is to ensure that data are scientifically sound
and of known precision and accuracy. This subsection (a)(2) discusses the 6
general categories of quality control measurements. Specific guidance is
provided for each method in the relevant protocol.
A) Calibration
Measurements. Calibration measurements are samples collected or measurements
made in a known radon environment, such as a radon chamber. Instruments providing
immediate results, such as continuous working level and radon monitors, shall
be operated in a radon chamber to establish individual instrument calibration
factors.
i) Calibration
measurements must be conducted to determine and verify the conversion factors
used to derive the concentration results. These factors are determined
normally for a range of concentrations and exposure times, and for a range of
other exposure and/or analysis conditions pertinent to the particular device.
ii) Determination
of these calibration factors is a necessary part of the laboratory analysis and
is the responsibility of the laboratory. These calibration measurement
procedures, including the frequency of tests and the number of devices to be
tested, shall be specified in the QAP maintained by manufacturers and analysis
laboratories.
iii) Licensees
providing measurements with active devices are required to recalibrate their
instruments at least once every 12 months.
B) Known
Exposure Measurements (Spikes). Known exposure measurements or spiked samples
consist of detectors that have been exposed to known concentrations in a radon
chamber. These detectors, such as charcoal canisters, alpha track detectors
and electret ion chambers, are labeled and submitted to the laboratory in the
same manner as ordinary samples to preclude special processing.
i) Suppliers
and analysis laboratories shall provide for the blind introduction of spiked
samples into their measurement processes and the monitoring of the results in
their QAP.
ii) Licensees
using passive monitors shall conduct spiked measurements (i.e., exposure in a
radon chamber where the environmental radon level is controlled) to aid the
Agency in verifying the accuracy of the entire measurement system. The licensee
shall conduct 3 spiked measurements per 100 measurements, with a minimum of 3
spiked measurements per year. For example, a licensee conducting only 70
measurements in a year must conduct at least 3 spiked measurements. A licensee
conducting 500 measurements during a one-year period must conduct at least 15
spiked measurements (3 per each 100). No more than 6 spiked measurements
will be required to
be taken within any single
month. For example, a licensee performing more than 200 measurements in one
month is not required to perform more than 6 spiked measurements that month.
Licensees are encouraged to take their spiked measurements from multiple
batches when possible and to take more than the minimally required spiked
measurements at their discretion. Devices shall be exposed in a radon chamber
at a minimum of 3 different radon concentrations, such as approximately 4.0,
10-30 and 30-100 pCi/L.
iii) Spikes shall be labeled in the same
manner as field detectors to ensure identical processing. The results of
analyses of detectors exposed to known radon concentrations shall be monitored
and recorded.
Any significant deviation from the known concentration to
which they were exposed shall be investigated and corrective action taken.
C) Background
Measurements. Background measurements are required both for continuous
monitors and for passive detectors requiring laboratory analysis.
i) Licensees
using continuous monitors shall perform sufficient instrument background
measurements to establish a reliable instrument background and to act as a
check on instrument operation.
AGENCY NOTE: Calibration
laboratories routinely perform background measurements of continuous monitors
during the calibration of instruments.
ii) Passive
detectors requiring laboratory analysis require one type of background
measurement made in the laboratory and another in the field.
iii) Laboratories
shall measure the background of a statistically significant number of unexposed
detectors from each batch or lot to establish the laboratory background for the
batch and the entire measurement system. This laboratory blank value is
subtracted (by the laboratory) from the field sample results reported to the
user, and shall be made available to the users for quality assurance purposes.
iv) Laboratories
performing these measurements shall calculate the lower limit of detection
(LLD) for their measurement systems. This LLD is based on the detector and
analysis system's background and can restrict the ability of some measurement
systems to measure low concentrations.
v) Licensees
using passive detectors shall employ field controls (called blanks) equal to
approximately 5 percent of the detectors that are deployed, or 25 each month,
whichever is smaller.
vi) These
controls shall be set aside from each detector shipment, kept sealed and in a
low radon environment, labeled in the same manner as the field samples to
preclude special processing, and returned to the analysis laboratory along with
each shipment. These field blanks measure the background exposure that may
accumulate during shipment and storage. The results shall be monitored and
recorded.
vii) The
recommended action to be taken if the concentrations measured by one or more of
the field blanks is significantly greater than the LLD is dependent upon the
type of detector and is discussed in the protocol for each method.
D) Duplicate
Measurements. Duplicate measurements provide a check on the precision of the
measurement result and allow the user to make an estimate of the relative
precision. Large precision errors may be caused by detector manufacture or
improper data transcription or handling by suppliers, laboratories, or
technicians
performing placements. Precision error can be an important
component of the overall error; therefore, licensees performing measurements
shall monitor precision.
i) Duplicate
measurements shall be side-by-side measurements made in at least 10 percent of
the total number of measurement locations, or 50 each month, whichever is
smaller. The locations selected for duplicate measurement shall be distributed
systematically throughout the entire population of samples.
ii) The
precision of duplicate measurements shall be monitored and recorded in the
quality assurance records. The analysis of data from duplicates shall be
plotted on range control charts. If the precision estimated by the user is not
within the precision expected of the measurement method, the cause of the
problem shall be investigated.
iii) Detectors
shall be treated identically in every respect. They shall be shipped, stored,
opened, installed, removed and processed together, and not identified as
duplicates to the processing laboratory.
E) Routine
Instrument Performance Checks. Proper functioning of analysis equipment and
operator usage require that the equipment and measurement system be subject to
routine checks. Regular monitoring of equipment and operators is vital to
ensure consistently accurate results. Performance checks include the frequent
use of an instrument check source. Components of the device (such as a pump,
battery or electronics) shall be checked regularly and the results noted in a
record. Each user shall develop methods for regularly monitoring (preferably
daily with use) their measurement system and for recording and reviewing
results.
F) Cross-checks.
Professional licensees using active monitors shall check their monitors for
bias on a regular basis. Ideally, such measurements are made in a radon
chamber. Exposure in a radon chamber is required during calibration. It can
be difficult to expose active monitors more often than once every 12 months.
It is important to more frequently assess the continued satisfactory operation
of the instrument response and to ensure damage from shipping has not occurred
prior to an instrument being placed into service after calibration.
Cross-checks shall be performed prior to placing an instrument being returned
to service after calibration and at 6 months (plus or minus a month) after
calibration. The following conditions shall be met:
i) Where
feasible, a cross-check shall begin with an instrument background measurement.
ii) The
cross-check measurement shall be made in an environment that has been chosen
for its stability and radon concentration that is above the lower limit of
detection.
iii) Cross-checks shall
be side-by-side measurements.
iv) One
of the instruments shall have been calibrated within the last 45 days.
v) A
measurement of at least 48 hours duration shall be conducted.
vi) The
bias of cross-check measurements shall be monitored and recorded in the quality
assurance records. If the bias estimated by the user is not within the bias
expected of the measurement, the cause of the problem shall be investigated and
corrective action taken in accordance with the licensee's Agency-approved QAP.
b) Protocol for using continuous radon monitors (CRs) to measure
indoor radon concentrations
1) Refer to Section 422.130 for a list of general conditions that
shall be met and standard information that shall be documented.
2)
When performing a radon
measurement, the CR shall be programmed to run continuously, recording
periodically (hourly or more frequently) the radon concentration for at least
48 hours. Longer measurements may be required per the continuous monitor type
and the radon level being measured.
3)
If the first 4 hours of
data from a 48-hour measurement are discarded because data are produced prior
to the establishment of equilibrium conditions in the test device, the
remaining hours of data shall be averaged and shall be sufficient to represent
a 2-day measurement.
4)
Every CR shall be
calibrated in a radon chamber, approved by the Agency, before being placed into
service, and after any repairs or modifications that could affect the
calibration. Subsequent recalibrations and background checks shall be
performed at least once every 12 months. Each scintillation cell requires an
individual calibration factor.
5) Background measurements shall be performed after every 1,000
hours of operation of scintillation cell-type CRs and whenever any type of CR
is calibrated. The background shall be checked by purging the monitor with
clean, aged air or nitrogen in accordance with the manufacturer's
instructions. In addition, the background count rate shall be monitored in
accordance with the manufacturer's instruction.
6)
Licensees providing
measurement services with CR devices shall perform duplicate measurements. The
performance and analysis of duplicates shall be completed in accordance with
subsection (a)(2)(D).
7) Pumps and flow meters shall be checked
before and after each measurement in accordance with the manufacturer's
instruction.
8) Licensees providing measurement
services with CR devices shall perform cross-checks. The performance and
analysis of cross-checks shall be completed in accordance with subsection
(a)(2)(F).
c) Protocol for using alpha track (AT) detectors to measure
indoor radon concentrations
1) Refer to Section 422.130 for a list of general conditions that
shall be met and standard information that shall be documented.
2) The laboratory background level for each batch of ATs shall be
established by each laboratory licensed by the Agency. Laboratories shall
measure the background of a statistically significant number of unexposed ATs
that have been processed according to the licensee's Quality Assurance Program
implementing/operating procedures.
3) Every AT laboratory system shall be calibrated in a radon
chamber at least once every 12 months. Determination of a calibration factor
requires exposures of ATs to a known radon concentration in a radon chamber.
These calibration exposures shall be used to obtain or verify the conversion
factor between net tracks per unit area and radon concentration.
A) ATs shall be exposed in a radon chamber at a minimum of 3
different radon concentrations such as approximately 4.0, 10-30 and 30-100
pCi/L or exposure levels similar to those found in the tested buildings.
B) Expose a minimum of 10 detectors at each radon concentration of
the chamber.
C) A calibration factor shall be determined for each batch or
sheet of detector material received from the supplier. Alternatively,
calibration factors may be established for several sheets, and these factors
extended to detectors from sheets exhibiting similar sensitivities (within
pre-established tolerance limits).
D) Analysis instruments shall be checked
at least daily for operability prior to operation. Analysis instruments do not
need to be checked on days not used.
4)
Licensees providing
measurement services with AT devices shall perform known exposure measurements
(spikes). The performance and analysis of spikes shall be completed in
accordance with subsection (a)(2)(B)
.
5)
Licensees providing
measurement services with AT devices shall perform duplicate
m
easurements. The performance and analysis of
duplicates shall be completed in accordance with subsection (a)(2)(D).
6) Licensees providing measurement
services with AT devices shall perform background measurements. The
performance of background measurements shall be completed in accordance with
subsection (a)(2)(C).
A) The results shall be monitored and
recorded. If one or a few field blanks have concentrations significantly
greater than the LLD established by the supplier, it may indicate defective
packaging or handling and the licensee shall investigate the cause. If the
average value from the field control devices (field blanks) is significantly
greater than the LLD established by the supplier, this average value shall be
subtracted from the individual values reported for the other devices in the
exposure group.
B) It may
be advisable to use 3 sets of detectors (pre-exposure, field and post-exposure
background) in order to allow the most thorough and complete evaluation of
radon levels. For example, one group of detectors (pre-exposure detectors) may
be earmarked for background measurement and returned for processing immediately
after the other detectors are deployed. The results from these detectors
determine if the number of tracks acquired before deployment is significant and
should be subtracted from the gross result. The second set of background
detectors (post-exposure background detectors) are obtained just before the
field monitors are to be collected and are opened and kept in the same location
as the returning field monitors for the same duration, and returned with them.
Finally, this "post-exposure background" is subtracted from the field
results, if found to be significant. In general, a value of 1 pCi/L or greater
for any blank AT indicates a significant level that should be investigated and
potentially subtracted from the field AT results.
d) Protocol for using electret ion chamber radon (ES or EL) detectors
to measure indoor radon concentration.
1) Refer to Section 422.130 for a list of general conditions that
shall be met and standard information that shall be documented.
2) Every short-term and long-term electret system and the
electret reader(s) shall be calibrated in a radon chamber, approved by the Agency.
Initial calibration for the system is provided by the manufacturer. Subsequent
recalibrations shall be performed at least once every 12 months. Determination
of calibration factors for short-term or long-term detectors requires exposure
of detectors to known concentrations of radon-222 in a radon exposure chamber.
Since short-term and long-term electret detector systems are also sensitive to
gamma radiation, a gamma exposure rate measurement in the test chamber is also
required annually.
3)
The following is provided
to manufacturers and suppliers of ES or EL services as minimum requirements in
determining the calibration factor:
A) Detectors shall be exposed in a radon chamber at a minimum of 3
different radon concentrations, such as approximately 4.0, 10-30 and 30-100
pCi/L, or exposure levels similar to those found in the tested buildings.
B) Expose a minimum of 10 detectors at each
radon concentration of the chamber.
C) Ensure a period of exposure sufficient
to allow the detector to achieve equilibrium with the radon chamber atmosphere.
4) Licensees providing measurement
services with ES or EL devices shall perform known exposure measurements (spikes).
The performance and analysis of spikes shall be completed in accordance with
subsection (a)(2)(B).
5) Licensees providing measurement
services with ES or EL devices shall perform duplicate measurements. The
performance and analysis of duplicates shall be completed in accordance with
subsection (a)(2)(D).
6) Licensees providing measurement services with short-term or
long-term electrets shall set aside a minimum of 5 percent of the electrets or
10, whichever number is smaller, from each shipment and evaluate them for
voltage drift. The electrets shall be kept covered with protective caps in a
low radon environment and analyzed for voltage drift over a time period similar
to the time period used for those deployed in measurements. Any voltage loss
found in the control electrets of more than one volt per week over a 3-week
test period for short-term electrets, or one volt per month over a 3-month
period for long-term electrets, shall be investigated.
7) Proper operation of the surface voltmeter shall be monitored
following the manufacturer's procedures for zeroing the voltmeter and analyzing
a reference electret. These checks shall be conducted at least once a week
while the voltmeter is in use.
8) All Laboratory Analysis licensees providing recharging
services of short-term or long-term electrets shall only provide those services
for devices they manufacture or for devices for which they have written
authorization from the manufacturer.
e) Protocol for using activated charcoal adsorption (AC) devices
to measure indoor radon concentrations
1) Refer to Section 422.130 for a list of general conditions that
shall be met and standard information that shall be documented.
2) Every activated charcoal adsorption system shall be calibrated
in a radon chamber at least once every 12 months. Determination of calibration
factors for ACs requires exposure of the detectors to known concentrations of
radon-222 in a radon chamber. The calibration factors depend on the exposure
time and may also depend on the amount of water adsorbed by the charcoal
container during exposure. Calibration factors shall be determined for each AC
measurement system (container type, amount of charcoal, gamma detector type,
etc.).
3)
Licensees providing
measurement services with AC devices shall perform known exposure measurements
(spikes). The performance and analysis of spikes shall be completed in
accordance with subsection (a)(2)(B).
4)
Licensees providing
measurement services with AC devices shall perform duplicate measurements. The
performance and analysis of duplicates shall be completed in accordance with
subsection (a)(2)(D).
5) Laboratory Control Detectors. The
laboratory background level for each batch of ACs shall be established by each
laboratory or supplier. Suppliers shall measure the background of a
statistically significant number of unexposed detectors that have been
processed according to their standard operating procedures (laboratory
blanks). The analysis laboratory or supplier calculates the net readings, that
are used to calculate the reported sample radon concentrations, by subtracting
the laboratory blank values from the results obtained from the field detectors.
6) Licensees providing measurement
services with AC devices shall perform background measurements. The
performance of background measurements shall be completed in accordance with
subsection (a)(2)(C).
A) One or a few of the field blanks have
concentrations significantly greater than LLD established by the supplier may
indicate defective devices or poor procedures and the licensee shall
investigate the cause.
B) If most of the field blanks have
concentrations significantly greater than the LLD, the average value of the
field blanks shall be subtracted from the reported field detector concentrations
and the supplier notified of a possible problem.
7) Counting equipment shall be subject to
daily operability checks by counting an instrument check source and determining
whether the reference source is constant to within established limits (2 standard
deviations). Daily operability checks do not need to be performed on days the
instrument is not used. The characteristics of the check source (geometry,
type of radiation emitted, etc.) shall be similar to those of the samples
analyzed. The count rate of the check sources shall be high enough to yield
good counting statistics in a short time (for example, 1000 to 10,000 counts
per minute) to provide a maximum random uncertainty of 5 percent.
f) Protocol for using charcoal liquid scintillation (LS) devices
to measure indoor radon concentrations
1) Refer to Section 422.130 for a list of general conditions that
shall be met and standard information that shall be documented.
2) Every LS laboratory system shall be calibrated in a radon
chamber at least once every 12 months. Determination of calibration factors
for LS devices requires exposure of calibration devices to known concentrations
of radon-222 in a radon chamber at carefully measured radon concentrations.
The calibration factors depend on the exposure time and may also depend on the
amount of water adsorbed by the device during exposure. Calibration factors
shall be determined for a range of different exposure times and, as
appropriate, humidities.
3)
Licensees providing
measurement services with LS devices shall perform known exposure measurements
(spikes). The performance and analysis of spikes shall be completed in
accordance with subsection (a)(2)(B).
4)
Licensees providing
measurements services with LS devices shall perform duplicate measurements.
The performance and analysis of duplicates shall be completed in accordance
with subsection (a)(2)(D).
5) Laboratory Control Devices. The
laboratory background level for each batch of LS devices shall be established
by each laboratory or supplier. Suppliers shall measure the background of a
statistically significant number of unexposed LS devices that have been
processed according to their standard operating procedures (laboratory
blanks). The analysis laboratory or supplier calculates the net readings that
are used to calculate the reported sample radon concentrations, by subtracting
the laboratory blank values from the results obtained from the field detectors.
6) Licensees providing measurement
services with LS devices shall perform background measurements. The
performance of background measurements shall be completed in accordance with
subsection (a)(2)(C).
A) One or a few of the field blanks have
concentrations significantly greater than the LLD established by the supplier
may indicate defective devices or poor procedures and the licensee shall
investigate the cause.
B) If most of the field blanks have
concentrations significantly greater than the LLD, the average value of the
field blanks shall be subtracted from the reported field detector
concentrations and the supplier notified of a possible problem.
7) Counting equipment shall be subject to
daily operability checks by counting an instrument check source and determining
whether the reference source is constant to within established limits (2 standard
deviations). Daily operability checks do not need to be performed on days the
instrument is not used. The characteristics of the check source (geometry,
type of radiation emitted, etc.) shall be similar to those of the samples
analyzed. The count rate of the check sources shall be high enough to yield
good counting statistics in a short time (for example, 1000 to 10,000 counts
per minute) to provide a maximum random uncertainty of 5 percent.
g) Protocol
for using continuous working level (CW) monitors to measure indoor radon progeny
concentrations
1) Radon Decay Product measurements may
be appropriate under certain conditions in large buildings, but are not
currently routinely performed by licensees or recommended by the American
Association of Radon Scientists and Technologists. The Agency does not
recommend their use for home environment or residential real estate
measurements. Licensees interested in using CWs for measurement purposes shall
submit Standard Operating Procedures, consistent with this Part, specific to
the model and design of the CW instrument to the Agency for approval.
2) Conditions and information in Section
422.130 shall be met.
3) Any measurement result based on radon
progeny shall be reported to no more than 3 decimal places, e.g., 0.033 working
level (WL).
4) The integrated average WL over the
measurement period shall be reported as the measurement result.
5) When performing a radon measurement,
the CW shall be programmed to run continuously, recording the periodic WL and,
when possible, the total integrated average WL. The longer the operating time,
the smaller the uncertainty associated with using the measurement result to
estimate a longer-term average concentration.
6) Working level values shall be
converted to pCi/L and both shall be reported to the client. The conversions
from WL to pCi/L shall be presented and explained clearly in the report to the
client. A statement shall be included in the measurement report that this
approximate conversion is based on a 40 percent equilibrium ratio. In
addition, the report shall state that this equilibrium ratio is typical, but
that any indoor environment may have a different and varying relationship
between radon and radon progeny.
7) Every continuous WL monitor shall be
calibrated in a radon chamber, approved by the Agency, before being placed into
service and after any repairs or modifications that could affect the
calibration. Subsequent recalibrations shall be performed at least once every
12 months.
8) Background measurements shall be
performed after every 168 hours of operation and whenever the unit is
calibrated. The CW shall be purged with clean, aged air or nitrogen in
accordance with the manufacturer's instructions. In addition, the background
count rate may be monitored more frequently by operating the CW in a low radon
concentration.
9) Measurement licensees providing measurement
services with CW devices shall perform duplicate measurements. The performance
and analysis of duplicates shall be completed in accordance with subsection
(a)(2)(D).
10) Pumps and flow meters shall be checked
before and after each measurement in accordance with the manufacturer's
instruction to ensure accuracy of volume measurements. This may be performed
using a dry-gas meter or other flow measurement device of traceable accuracy.
11) Licensees providing measurement
services with CW devices shall perform cross-checks. The performance and
analysis of cross-checks shall be completed in accordance with subsection
(a)(2)(F).