77 Ill. Adm. Code 250.1090
Sterilization and Processing of Supplies
Section 250
Section 250.1090Â
Sterilization and Processing of Supplies
a)Â Â Â Â Â Â Â Â All sterilization and processing of all sterile supplies and
equipment shall be under competent, qualified supervision.
1)Â Â Â Â Â Â Â Â The director or person responsible for central services shall
be responsible to the chief executive officer either directly or through a
designated department head. The director of the central sterile supply shall
be qualified for the position by education, training, and experience and shall
be a member of the Infection Control Committee. (See Section 250.1100(a).)
2)Â Â Â Â Â Â Â Â The number of supervisory and support personnel shall be
related to the scope of the services provided. New employees shall receive
initial orientation and on-the-job training, and all employees shall
participate in a continuing in-service education program, which shall be
documented.
3)Â Â Â Â Â Â Â Â Educational efforts, though directed primarily at
sterile-supply processing and handling techniques, shall also include
management concepts, safety, personal hygiene, health requirements, and work
attire.
b)Â Â Â Â Â Â Â Â There shall be written policies and procedures for the
decontamination and sterilization activities performed in central services and
elsewhere in the hospital. The hospital shall comply with the Centers for
Disease Control and Prevention Guidelines for Disinfection and Sterilization in
Healthcare Facilities. These policies and procedures shall include, but are
not limited to, the following:
1)Â Â Â Â Â Â Â Â The receiving, decontaminating, cleaning, preparing,
disinfecting and sterilizing of reusable items.
2)Â Â Â Â Â Â Â Â The assembly, wrapping, storage, distribution, and quality
control of sterile equipment and medical supplies. Load control numbers shall
be used to designate the hospital sterilization equipment used for each item,
including the sterilization date and cycle.
3)Â Â Â Â Â Â Â Â The use of sterilization process monitors, including
temperature and pressure recordings, and the use and frequency of appropriate
chemical indicator and bacteriological spore tests for all sterilizers.
4)Â Â Â Â Â Â Â Â Designation of the shelf life for each hospital-wrapped and
-sterilized medical item and, to the maximum degree possible, for each
commercially prepared item.
A)Â Â Â Â Â Â Â Designation of a shelf life may be a specific expiration date,
i.e., 30 days, six months, etc., based on manufacturer's recommendation, a
nationally recognized authority, or other standard approved by the facility's
Infection Control Committee.
B)Â Â Â Â Â Â Â Designation of shelf life may be event related if policies and
procedures, approved by the Infection Control Committee, address at least the
following:
i)Â Â Â Â Â Â Â Â Â requirements for wrapping, storage and rotation of sterile
supplies;
ii)Â Â Â Â Â Â Â Â definition of an event that may cause a sterile item to be or
be suspected of being compromised, such as the package being wet or torn, or
the seal being broken or tampered with;
iii)Â Â Â Â Â Â Â clear direction that the final inspection of the package and
the ultimate decision to use the contents of the package rest with the
clinician; and
iv)Â Â Â Â Â Â Â orientation, in-service and other follow-up training to assure
that all necessary staff understand and implement the policies and procedures.
C)Â Â Â Â Â Â Â A facility may choose to use both a specific expiration date
and event-related shelf life designation specific for certain wrappings, areas
of the hospital, etc., as long as the policies and procedures, as approved by
the Infection Control Committee, and the training of staff define this
practice.
5)Â Â Â Â Â Â Â Â Acquisition of supplies after normal working hours or any time
the central service or sterile supply unit is considered "closed" or
unstaffed.
6)Â Â Â Â Â Â Â Â Preventive maintenance of all central supply service
equipment, including performance verification records and reports.
7)Â Â Â Â Â Â Â Â The recall and disposal or reprocessing of expired or
inadequately sterilized supplies.
8)Â Â Â Â Â Â Â Â The emergency collection and disposition of supplies when
special warnings have been issued by the manufacturer. The attending physician
shall be notified when patient exposure is known.
9)Â Â Â Â Â Â Â Â Specific aeration requirements for each category of
gas-sterilized items to eliminate the hazard of toxic residues.
10)Â Â Â Â Â Â Â Â The cleaning and sanitizing of work surfaces, floors,
utensils, and equipment used in central service functions.
c)Â Â Â Â Â Â Â Â Space shall be provided for the efficient operation of all
central service functions. Functional design and work flow patterns shall
provide for the separation of soiled and contaminated supplies from those that
are clean and sterile. Equipment of adequate design, size, and type shall be
provided for the effective decontaminating, disinfecting, cleaning, packaging,
sterilizing, storing, and distributing of medical instruments, supplies, and
equipment used in patient care.
d)Â Â Â Â Â Â Â Â Equipment and procedures
1)Â Â Â Â Â Â Â Â The facilities, equipment, and procedures for clean-up,
preparation, and sterilization shall be adequate to allow proper cleaning,
processing, and sterilizing of patient care supplies and equipment.
2)Â Â Â Â Â Â Â Â When clean-up, preparation, and sterilization functions are
carried out in the same room or unit (as in a central sterilizing department)
the physical facilities and equipment and the policies and procedures for their
use shall be such as to effectively separate soiled or contaminated supplies
and equipment from the clean or sterilized supplies and equipment.
3)Â Â Â Â Â Â Â Â Sterilization equipment shall be maintained in good repair and
under the provisions of a preventive maintenance program of the Engineering and
Maintenance Services. (Refer to Subpart P.)
4)Â Â Â Â Â Â Â Â All pressure steam autoclaves shall have recording
thermometers, and the sterilization performance shall be otherwise checked.
e)Â Â Â Â Â Â Â Â Sterilization of instruments and utensils
1)Â Â Â Â Â Â Â Â All surgical instruments not adversely affected by high
temperature shall be sterilized by pressure steam sterilization.
2)Â Â Â Â Â Â Â Â The steam method of sterilization is the preferred method for
sterilizing medical and surgical instruments that are not damaged by heat,
steam, pressure, or moisture. Low-temperature sterilization technologies
(e.g., Ethylene Oxide (EtO), hydrogen peroxide gas plasma) may be used for
reprocessing patient care equipment that is heat or moisture sensitive. In
addition, a peracetic acid immersion system of sterilization may be used to
sterilize heat-sensitive immersible medical and surgical items, and dry-heat
sterilization may be used to sterilize items (e.g., powders, oils) that can
sustain high temperatures. Operating parameters and guidelines for each method
or system of sterilization shall be followed for whichever method is used.
3)Â Â Â Â Â Â Â Â All instruments shall be thoroughly cleaned before
sterilization.
4)Â Â Â Â Â Â Â Â Boiling is not an approved method of sterilization.
f)Â Â Â Â Â Â Â Â Water sterilization
1)Â Â Â Â Â Â Â Â When non-commercial sterile water is utilized, water
sterilization equipment shall be maintained and operated in a manner that will
protect the sterilized water from contamination.
2)Â Â Â Â Â Â Â Â An acceptable method for checking the sterility of the water
shall be utilized. Water may be sterilized either in approved water sterilizers
or autoclaved in approved flasks.
g)Â Â Â Â Â Â Â Â Sterilization and storage of supplies and equipment
1)Â Â Â Â Â Â Â Â Supplies and equipment shall be properly wrapped and labeled
before sterilization.
2)Â Â Â Â Â Â Â Â The effectiveness of hospital sterilization shall be checked.Â
Mechanical, chemical, and biologic monitors shall be used to ensure the
effectiveness of the sterilization process.  Indicators shall be used to show
that the items have been sterilized. A procedure shall be established for the
recall of expired or inadequately sterilized goods for both in-house and
commercially sterilized supplies and equipment. Refer to Section 250.1100(a).
3)Â Â Â Â Â Â Â Â Supplies and equipment commercially prepared so as to retain
sterility indefinitely are acceptable. The hospital shall satisfy itself of
the sterility of such materials.
4)Â Â Â Â Â Â Â Â Sterile equipment and supplies shall be stored properly in clean
cabinets, cupboards or other suitable enclosed spaces. An orderly system of
rotation of supplies is recommended so that supplies stored first will be used
first.
h)Â Â Â Â Â Â Â Â Transmissible spongiform
encephalopathies (TSEs)
1)Â Â Â Â Â Â Â Â Records
shall be maintained for at least 20 years regarding quarantine, disposal,
decontamination, and sterilization of surgical instruments used for patients
with a confirmed or suspected TSE.
2)Â Â Â Â Â Â Â Â For
the purposes of this Section, TSEs are a group of rapidly progressive, invariably
fatal neurodegenerative diseases that affect both humans and animals. TSEs in
humans include Creutzfeldt-Jakob disease (CJD), kuru,
Gerstmann-Straussler-Scheinker syndrome (GSS), fatal familial insomnia (FFI),
and variant CJD (vCJD).