77 Ill. Adm. Code 518.1950
Sterilization and Processing of Supplies
Section 518
Section 518.1950Â
Sterilization and Processing of Supplies
a)Â Â Â Â Â Â Â Â All sterilization and processing of all sterile supplies and
equipment shall be under competent, qualified supervision.
1)Â Â Â Â Â Â Â Â The director or person responsible for sterile supplies and
equipment shall be responsible to the chief executive officer. This person
shall be qualified for the position by education, training and experience.
2)Â Â Â Â Â Â Â Â The number of supervisory and support personnel shall be
related to the scope of the services provided. New employees shall receive
initial orientation and on-the-job training, and all employees shall
participate in a continuing in-service education program, which shall be
documented.
3)Â Â Â Â Â Â Â Â Educational efforts, though directed primarily at
sterile-supply processing and handling techniques, shall also include
management concepts, safety, personal hygiene, health requirements and hand-washing,
and work attire.
b)Â Â Â Â Â Â Â Â Written policies and procedures shall be established for the
decontamination and sterilization activities performed in the freestanding
emergency center and shall relate, but are not limited, to the following:
1)Â Â Â Â Â Â Â Â Receiving, decontaminating, cleaning, preparing, disinfecting
and sterilizing reusable items.
2)Â Â Â Â Â Â Â Â Assembly, wrapping, storage, distribution, and quality control
of sterile equipment and medical supplies. Â Load control numbers shall be used
to designate the sterilization equipment used for each item, including the
sterilization date and cycle.
3)Â Â Â Â Â Â Â Â Use of sterilization process monitors, including temperature
and pressure recordings, and the use and frequency of appropriate chemical
indicator and bacteriological spore tests for all sterilizers.
4)Â Â Â Â Â Â Â Â Designation of the shelf life for each FEC-wrapped and
-sterilized medical item and, to the maximum degree possible, for each
commercially prepared item.
A)Â Â Â Â Â Â Â Designation of a shelf life may be a specific expiration date,
i.e., 30 days, six months, etc., based on manufacturer's recommendation, a
nationally recognized authority, or other standard approved by the owning or
controlling hospital's Infection Control Committee.
B)Â Â Â Â Â Â Â Designation of shelf life may be event related if policies and
procedures, approved by the owning or controlling hospital's Infection Control
Committee, address at least the following:
i)Â Â Â Â Â Â Â Â Â Requirements for wrapping, storing and rotating sterile
supplies;
ii)Â Â Â Â Â Â Â Â Definition of an event that may cause a sterile item to be or
be suspected of being compromised, such as the package being wet or torn, or
the seal being broken or tampered with;
iii)Â Â Â Â Â Â Â Clear direction that final inspection of the package and the
ultimate decision to use the contents of the package rest with the clinician;
and
iv)Â Â Â Â Â Â Â Orientation, in-service and other follow-up to assure that all
necessary staff understand and implement the policies and procedures.
C)Â Â Â Â Â Â Â A facility may choose to use both a specific expiration date
and event-related shelf life designation specific for certain wrappings, areas
of the FEC, etc., as long as the policies and procedures, as approved by the
Infection Control Committee, and training of staff define this practice.
5)Â Â Â Â Â Â Â Â Acquisition of supplies after normal working hours or any time
the central supply service or sterile supply unit is considered
"closed" or unstaffed.
6)Â Â Â Â Â Â Â Â Preventive maintenance of all central supply service
equipment, including performance verification records and reports.
7)Â Â Â Â Â Â Â Â The recall and disposal or reprocessing of outdated sterile
supplies.
8)Â Â Â Â Â Â Â Â The emergency collection and disposition of supplies when
special warnings have been issued by the manufacturer. The attending physician
shall be notified if patient exposure is known.
9)Â Â Â Â Â Â Â Â Specific aeration requirements for each category of
gas-sterilized items to eliminate the hazard of toxic residues.
10)Â Â Â Â Â Â Â The cleaning and sanitizing of work surfaces, floors, utensils
and equipment used in central supply service functions.
c)Â Â Â Â Â Â Â Â Space shall be provided for the efficient operation of all
central supply service functions. Functional design and work-flow patterns
shall separate soiled and contaminated supplies from supplies that are clean
and sterile. Equipment of adequate design, size and type shall be provided for
decontaminating, disinfecting, cleaning, packaging, sterilizing, storing and
distributing medical instruments, supplies and equipment used in patient care.
d)Â Â Â Â Â Â Â Â Equipment and Procedures
1)Â Â Â Â Â Â Â Â The facilities, equipment, and procedures for cleanup,
preparation, and sterilization shall be adequate to allow proper cleaning,
processing, and sterilizing of patient care supplies and equipment.
2)Â Â Â Â Â Â Â Â When cleanup, preparation, and sterilization functions are
carried out in the same room or unit (as in a central sterilizing department),
the physical facilities and equipment and the policies and procedures for their
use shall separate soiled or contaminated supplies and equipment from the clean
or sterilized supplies and equipment.
3)Â Â Â Â Â Â Â Â Sterilization equipment shall be maintained in good repair and
be under a preventive maintenance program.
4)Â Â Â Â Â Â Â Â All pressure steam autoclaves shall have recording
thermometers, and the sterilization performance shall be otherwise monitored.
e)Â Â Â Â Â Â Â Â Sterilization of Instruments and Utensils
1)Â Â Â Â Â Â Â Â All surgical instruments not adversely affected by high
temperature shall be sterilized by pressure steam sterilization.
2)Â Â Â Â Â Â Â Â Whenever possible, throughout the FEC, sterilization shall be
accomplished by pressure steam sterilization. Hot air sterilization or gas
sterilization may be used. When gas sterilization is used, there shall be
policies and tested procedures for proper aeration to permit safe use. Pressure
steam sterilization of reusable syringes and needles is required.
3)Â Â Â Â Â Â Â Â All instruments, whether used on infected cases or clean
cases, shall be cleaned before sterilization. Instruments used on infected
cases shall be disinfected before transport to central supply.
4)Â Â Â Â Â Â Â Â Boiling is not an approved method of sterilization.
f)Â Â Â Â Â Â Â Â Water Sterilization
1)Â Â Â Â Â Â Â Â When non-commercial sterile water is used, water sterilization
equipment shall be maintained and operated in a manner that will protect the
sterilized water from contamination.
2)Â Â Â Â Â Â Â Â An acceptable method for checking the sterility of the water
shall be used. Water may be sterilized either in water sterilizers or
autoclaved in appropriate flasks.
g)Â Â Â Â Â Â Â Â Sterilization and Storage of Supplies and Equipment
1)Â Â Â Â Â Â Â Â Supplies and equipment shall be properly wrapped and labeled
before sterilization.
2)Â Â Â Â Â Â Â Â The effectiveness of sterilization shall be checked. This shall
include bacteriological testing of all sterilization units throughout the
facility. Indicators shall be used to show that a wrapped package has been
sterilized. A procedure shall be established for the recall of expired or
inadequately sterilized goods for both in-house and commercially sterilized
supplies and equipment.
3)Â Â Â Â Â Â Â Â Supplies and equipment commercially prepared so as to retain
sterility indefinitely are acceptable. The FEC shall verify the sterility of these
materials.
4)Â Â Â Â Â Â Â Â Sterile equipment and supplies shall be stored properly in
clean cabinets, cupboards or other suitable enclosed spaces. An orderly system
of rotation of supplies is recommended so that supplies stored first will be
used first.