77 Ill. Adm. Code 682.170
Audiometer Calibrations
Section 682
Section 682.170 Audiometer
Calibrations
An annual calibration shall be
conducted on each audiometer used in dispensing hearing instruments.
a) Audiometer calibration data sheets shall be kept on file at
the licensed hearing instrument dispenser's/audiologist's place of business for
four years after the date of calibration.
b) The audiometer calibration data sheet shall include the
following:
1) Audiometer identification, consisting of make, model, and
serial number;
2) The calibrator's identification, consisting of the company
name, the company address, and the name of the individual who conducted the
calibration;
3) Audiometer calibration readings for air and bone conduction,
speech, rise and decay time, and masking;
4) The calibrator's certification that the audiometer meets or
exceeds American National Standards Institute (ANSI) standards (see Section
682.105(a)(1)); and
5) Date of calibration.
c) Calibration shall be accomplished by the manufacturer or a
person equipped with instruments for calibrating audiometers.
d) Calibration of audiometers shall be in accordance with the standards
set by ANSI (see Section 682.105(a)(1)).
e) The licensed hearing instrument dispenser/audiologist shall
indicate the make of the audiometer, the model, serial number, and the date of
the last ANSI calibration for each audiometer used in hearing instrument
dispensing activities on the Audiometer Calibration Form. The Form shall be
signed and shall be presented to the Department upon request.