310 CMR 42.08
Minimum Standards for Certification
(1) Personnel. Certified laboratories and laboratories seeking certification shall designate a
laboratory director and laboratory supervisor(s), and employ a sufficient number of qualified
analysts commensurate with the laboratory's workload. The designated laboratory director and
laboratory supervisor may be the same individual if he or she possesses the minimum
qualifications and fulfills the responsibilities of both positions as set in 310 CMR 42.08(1).
(a) The laboratory director shall have the following responsibilities:
1. Developing policies, programs, and standard operating procedures that will ensure
accurate and objective analytical results;
2. Employing and ensuring the training of qualified laboratory personnel;
3. Reporting analytical results to the Department in accordance with 310 CMR 42.13;
and
4. Interpreting and evaluating reports of analyses submitted by the laboratory upon
request of the Department.
(b) A laboratory supervisor shall have the following responsibilities:
1. Performing analyses and/or providing personal and direct supervision and training
to laboratory analysts performing analyses in the categories for which the supervisor is
qualified;
2. Instruction and general supervision of all other laboratory operations including
sample tracking, data validation, quality control, verification, and prompt reporting of
results; and
3. Acceptance and non-acceptance of samples submitted to the laboratory for analysis.
(c) Minimum Qualifications of Laboratory Director.
1. Academic Training. The laboratory director shall possess a Bachelor's degree in
biology, chemistry, or a closely related field. If chemical analyses are to be performed
by the laboratory, the director must have at least 24 college credits in chemistry.
2. Experience. The laboratory director shall have a minimum of three years of
experience in an environmental analysis laboratory.
(d) Minimum Qualifications of Laboratory Supervisor for Laboratories Certified in the
Discipline of Chemistry or Radiochemistry.
1. Academic Training. The laboratory supervisor shall possess a Bachelor's degree in
chemistry, biology, or a closely related field, and have at least 30 college credits in
chemistry.
2. Experience.
a. Chemistry. The laboratory supervisor shall have a minimum of two years of
laboratory experience in chemical analysis including one year of experience in the
specific chemical methods supervised.
b. Radiochemistry. The laboratory supervisor shall have a minimum of two years
of laboratory experience in radiochemical analyses, including one year of experience
in the radiochemical methods for which the laboratory is certified.
(e) Minimum Qualifications of Laboratory Supervisor for Laboratories Certified in the
Discipline of Microbiology.
1. Academic Training. The laboratory supervisor shall have a minimum of a Bachelor's
degree in biology, chemistry, or a closely related field with at least four college credits
in microbiology.
2. Experience. The laboratory supervisor shall have a minimum of one year of
experience in the microbiological methods for which the laboratory is certified.
(f) Minimum Qualifications of Instrumentation Analyst.
1. Academic Training. An instrumentation analyst shall possess a minimum of a high
school diploma or equivalent and eight college credits in chemistry.
2. Experience. An instrumentation analyst shall have a minimum of one year of training
or experience in the operation of the appropriate instrumentation.
(g) Minimum Qualifications of Analyst.
1. Academic Training. An analyst shall possess a minimum of a high school diploma
or equivalent.
2. Training. An analyst shall receive specialized training in the methods to be
performed.
(h) The Department may exempt a laboratory holding certification status on the effective
date of 310 CMR 42.00 from one or more of the requirements set forth in 310 CMR
42.08(1), if the Department finds that strict compliance with such requirements would result
in an undue hardship and would not serve to further the intent of 310 CMR 42.00. An
exemption, when granted, shall be effective for not more than one year unless renewed by
the Department.
(2) Facilities. A certified laboratory shall have:
(a) Adequate space in which to perform the analyses and related activities in the disciplines
and categories in which it is certified;
(b) A well-lighted laboratory work bench area of ample size that is convenient to a sink, hot
and cold running water, gas, suction, and electrical outlets if necessary;
(c) Clear aisles between benches that provide adequate room for passage of personnel and
equipment;
(d) Appropriate ventilation including exhaust hoods for the handling of chemicals and
samples in order to limit contamination of samples, standards, and other laboratory areas
with the performance of each exhaust hood and biological safety cabinet tested by a qualified
person annually and determined to be operating satisfactorily;
(e) Appropriate facilities for the storage of volatile, corrosive, and flammable materials;
(f) Controlled laboratory temperature and humidity at the levels required for the proper
performance of the analyses and for the optimum operation of instruments that are sensitive
to variations in temperature; and
(g) A locked facility for storage of chain-of-custody samples.
(3) Equipment and Materials. Certified laboratories and laboratories seeking certification shall
have readily available on the premises all equipment, supplies, reagents, glassware, and
instrumentation necessary to perform the analyses for which the laboratory is certified or seeks
certification and related quality control activities. Such equipment and materials shall be
maintained in good working condition, meet the performance criteria of the analytical method
used, and, where appropriate, meet the criteria specified in 310 CMR 42.08(3):
(a) Refrigerator. Aqueous reagents and samples may be stored in a standard household
refrigerator. A flammable materials refrigerator shall be used for storage of organics and
flammable materials. The internal temperature of a refrigerator shall be maintained at 4°C
+ 2°C. Organics and flammable materials not requiring refrigeration shall be stored in a
flammable storage cabinet.
(b) Drying Oven. The drying oven shall have selectable temperature control with a range
from room temperature to 180°C + 2°C or higher.
(c) Source of Distilled or Deionized Water. Distilled or deionized water shall meet the
minimum criteria of the methodologies employed.
(d) Top-loader Balance. The balance pan shall be clean and free of corrosion. The balance
must be capable of detecting a weight of 100 mg at a 150 g load or one mg for a load of 10 g
or less.
(e) Hot Plate. The hot plate shall have selectable temperature controls for safe heating of
laboratory reagents.
(f) Magnetic Stirrer. The magnetic stirrer shall be of variable speed with a stirring bar
coated with inert material.
(g) Glassware. Glassware shall be made of borosilicate glass. Volumetric glassware shall
be marked “Class A”.
(h) Analytical Balance. The analytical balance shall be readable to 0.1 mg. It shall be
mounted on a stable shock-proof base and protected from interference due to air currents.
(i) Conductivity Meter. The conductivity meter shall be readable in appropriate units, have
a range from 2 ohms to 2 megohms or equivalent micromhos or Siemens, and be capable of
measuring conductivity with an error not exceeding + 1% or one :S/cm, whichever is
greater.
(j) pH Meter. The pH meter shall be accurate to at least + 0.05 units and have scale
readability to + 0.01 units.
(k) Thermometer. The thermometer shall have 1°C or finer subdivisions and be calibrated
in degrees Celsius for the temperature range in which it will be used. The column shall have
no separations. An organic solvent-filled thermometer may be used in a refrigerator. The
laboratory may use electronic temperature measurement devices having the appropriate
accuracy and readability for their intended use.
(l) Water Baths. Water baths may be electric or steam heated and capable of controlling
temperature to 100°C within 5°C.
(m) Incubators. Incubators shall have an internal temperature-monitoring device and be
capable of maintaining proper temperature.
(n) Autoclave. The autoclave shall be equipped with an accurate thermometer, a separate
pressure gauge, and an operational safety valve. It shall maintain the sterilization
temperature throughout the sterilization cycle and depressurize slowly so that no air bubbles
form in inverted tubes and medium does not boil over. When being used to sterilize
carbohydrate-containing media, the autoclave must be capable of completing an entire cycle
(i.e., time when materials are exposed to heat) within 45 minutes when a 12-15 minute
sterilization period is used.
(o) Hot-air Sterilization Oven. Hot-air sterilization ovens shall be capable of maintaining
a stable sterilization temperature (170°C - 180°C).
(p) Muffle Furnace. Muffle furnaces shall be capable of heating glassware to 400°C for
cleaning.
(q) Reagents, Standards, Media. Consumable supplies such as, but not limited to, reagents,
standards, and media must not be used beyond their expiration date.
(r) Stereo Microscope. Stereo microscopes must have a magnification of 10-15x and be
equipped with a fluorescent light source.
(s) Colony Counter. When the pour plate technique is used, a dark-field colony counter
must be used to count heterotrophic plate count colonies.
(4) Laboratory Safety Measures Affecting Laboratory Analysis Capability. Certified
laboratories shall be in compliance with local, state, and federal laws to the extent that required
conditions directly affect laboratory capability in providing accurate and reliable analysis and
maintaining the integrity of samples and of the analytical process; or are referenced as required
for environmental laboratories by the acceptable analytical method(s) employed. These
conditions include, but are not limited to, safety protocols involving physical, chemical,
radiochemical, and biological hazards; availability and use of fire safety equipment; protective
clothing and equipment including respirators; appropriate handling of compressed gas cylinders;
and appropriate storage and disposal of materials and wastes.
(5) Quality Assurance/Quality Control.
(a) General Requirements.
1. Quality Assurance Plan. Certified laboratories shall establish, maintain, and follow
a written Quality Assurance (QA) Plan acceptable to the Department. Each laboratory's
QA plan shall be made available to all analysts employed by the laboratory. At a
minimum, QA plans shall include:
a. QA organization and responsibility;
b. QA objectives for precision and accuracy;
c. Standard operating procedures (SOPs) that accurately reflect all phases of current
laboratory activities. The standard operating procedures section of a laboratory's QA
plan shall include the following:
i. Sample receipt and handling procedures including sample custody and storage
procedures;
ii. Calibration procedures and frequencies;
iii. Analytical procedures, including any modifications to published procedures;
iv. Data reduction, validation and reporting procedures;
v. Internal quality control procedures (type and frequency);
vi. Provision for performance and system audits, both internal and external, and
schedules;
vii. Preventive maintenance procedures and schedules;
viii. Specific procedures for assessing data precision and accuracy;
ix. Procedures for taking corrective actions;
x. Quality assurance reporting procedures; and
xi. Laboratory safety plans.
A standard operating procedure for an analytical method or other laboratory
procedure may be a separate document. The QA plan must include a list of all such
standard operating procedures. Only the laboratory director or supervisor may make
changes in standard operating procedures. Such changes shall be effective only when
put in writing.
d. Record Maintenance. The record maintenance procedures section of a
laboratory's QA plan shall include the procedures for creating, controlling, and
maintaining the following records:
i. Raw data (including, but not limited to, laboratory notebooks, instrument
printouts, and electronic records);
ii. Chain-of-custody records;
iii. Calculations;
iv. Quality control data; and
v. Reports.
2. Temperature Records.
a. A laboratory shall measure and record the temperature of each drying oven and
hot-air sterilizing oven for each day of use. The temperature measurement device
must be immersed in sand or other suitable material and placed on one of the shelves.
A laboratory may use a temperature-measuring device that can be read from outside
the oven without the need to open the door provided that it has verified the accuracy
of the device.
b. A laboratory shall measure and record the temperature of each refrigerator and
incubator either continuously or each day of use. The thermometers must be
immersed in liquid and placed on one of the shelves. A laboratory may use a
temperature-measuring device that can be read from outside the refrigerator or
incubator without the need to open the door provided that it has verified the accuracy
of the device.
3. Laboratory Chemicals and Reagents. Analytical reagent (AR) grade or American
Chemical Society (ACS) grade chemicals or better are required for analyses, unless
otherwise required by the analytical method. In addition, laboratory chemicals and
reagents shall meet the following requirements:
a. All chemicals shall be labeled with the date of receipt by the laboratory to prevent
the use of outdated reagents;
b. Stock and working standard solutions shall be compared with check standards
and inspected prior to use for signs of decomposition, such as formation of
precipitates, evaporation, and/or discoloration;
c. All reagents and standards shall be labeled with identification of the compound,
concentration, solvent, date of preparation, date of expiration, and the name of the
analyst who prepared the solution;
d. Preparation of all stock standards and primary dilution standards shall be
documented, and the concentration of stock and working calibration standards shall
be verified against a primary dilution standard prepared from a source of reagents
different from those used to prepare the calibration standards; and
e. The use of the acceptable grade of reagents and compressed gases required by the
analytical procedure employed shall be documented.
4. Laboratory Glassware. The laboratory must follow cleaning procedures for glassware
and other labware that are specified in the analytical methods. If no specifications are
given in a method, then glassware and sample containers must be cleaned prior to use
by washing in a warm detergent solution, followed by thorough rinsing with tap water
and several additional rinses using deionized or distilled water. The laboratory must use
a detergent designed for laboratory use. Commercially prepared glassware and sample
containers may be used provided the laboratory documents the source and cleaning
procedures utilized. Certain analytical methods may require additional glassware
preparation procedures or the maintenance of a separate dedicated set of glassware.
5. Maintenance of Laboratory Instrumentation and Equipment. Analytical
instrumentation and equipment shall be maintained in accordance with the manufacturer's
instructions, analytical method requirements, and good laboratory practices. A secure
record of maintenance procedures shall be maintained for each instrument and piece of
equipment.
6. Instrumentation Calibration Requirements.
a. General Requirements.
i. Unless directed otherwise by the analytical method employed, all instruments
shall be calibrated immediately prior to analysis using a minimum of a blank and
three calibration standards that bracket the expected concentration range.
ii. Unless directed otherwise by the analytical method employed, all instrument
calibrations shall be verified through the analysis of a calibration check sample
that has been prepared using a source of reagents different from that used to
prepare the calibration standards. Unless directed otherwise by the analytical
method employed, the calibration check sample shall be analyzed at the
beginning and at the conclusion of the analysis session and after every 20 or
fewer samples. If the result does not agree within 20% of the original value,
corrective action shall be taken.
iii. For instruments with a calibration curve that has been set by the instrument
manufacturer, the laboratory shall verify the calibration curve using a minimum
of three calibration check standards that bracket the expected concentration
range. The check standards shall represent low, medium, and high concentrations
and include a standard at the minimum reporting level (MRL). If the result of the
calibration check does not agree within 10% of the assigned value of each check
standard, instrument recalibration must be performed.
iv. The laboratory shall keep a secure record of instrument calibration
procedures.
b. Analytical Balance. Each analytical balance shall be checked and adjusted
annually by a qualified service person. The accuracy of each analytical balance shall
be checked each day it is to be used using a minimum of two ASTM Class 1 weights,
or equivalent, in ranges appropriate to the laboratory’s weighing needs. The
laboratory shall keep a secure record of the results of accuracy checks, the date
performed, and the signature of the analyst who performed the check. The non-
reference weights used for this check must be calibrated annually using reference
weights and the results recorded. The accuracy of reference weights must be certified
every five years. The balance level shall be checked prior to each use and adjusted
if necessary.
c. pH Meters. Each pH meter shall be calibrated daily or prior to each use with
pH 7.0 and pH 4.0 or 10.0 buffer standards that bracket the expected value of the
sample, medium, or reagent being tested. The buffers used shall be recorded,
including the date of calibration and the name of the analyst who performed the
calibration.
i. The laboratory must use pH buffer aliquots only once.
ii. The laboratory must date commercial buffer solution containers upon receipt
and when opened.
iii. The laboratory must record the pH meter slope monthly after calibration and
take corrective action if the slope falls outside the range of 95% to 105%.
d. Conductivity Meters.
i. The conductivity meter must be calibrated at least monthly using a certified
and traceable low-level standard. Alternatively, the laboratory must determine
the cell constant monthly.
ii. An in-line meter may be used to check reagent-grade water provided that it
is calibrated annually.
e. Thermometers. The accuracy of all temperature measurement devices used to
monitor temperatures shall be verified by comparing the reading of each device with
that of a certified reference thermometer that is graduated in degree increments no
larger than those of the device whose accuracy is being verified. The laboratory must
discontinue use of a thermometer graduated in 0.5/C increments or less that differs
from the certified thermometer by more than 1/C. The accuracy of glass and
electronic thermometers must be verified annually; metal thermometers must be
verified quarterly; infrared detection devices must be verified every six months; and
the certified reference thermometer must be calibrated at least once every five years.
The correction factor and date of verification of accuracy must be indicated on the
thermometers. The laboratory shall maintain a secure record that includes:
i. The identification number of each thermometer;
ii. The temperatures displayed on both the certified thermometer and the
thermometer being verified;
iii. Any applicable correction factor;
iv. The date each check was performed; and
v. The signature of the analyst who performed each check.
f. On-line Monitors and Portable Equipment. Continuous on-line monitors and
portable equipment used in obtaining on-site measurements must be calibrated in
accordance with the manufacturer's instructions. The calibration must be verified
through analysis of an independent check sample or use of an independent
monitoring technique. Verification shall be recorded.
g. Spectrophotometer Wavelength Verification. The wavelength setting of the
spectrophotometer shall be checked annually by comparing the wavelength setting
to that of colored standards or filters, such as didymium glass. The wavelength
observed, date of performance, and the name of the analyst or service person that
performed the check shall be recorded.
h. Top-loader Balance. Each top-loader balance must be checked for accuracy
monthly in its range of use with ASTM Class 1 weights or equivalent.
7. Sample Collection, Preservation and Handling.
a. Acceptable procedures, as referenced or defined in current federal regulations
shall be utilized for sample collection, handling and preservation.
b. The laboratory may reject the sample if it is not assured of the sample
identification or of the validity of the sample collection, handling and preservation
procedures. The laboratory must have a written policy listing its criteria for rejection
of samples. When rejecting a specific sample, the laboratory must document the
reason(s) for the rejection.
c. Samples shall be stored in such a way that cross-contamination from other
samples, standards or reagents is avoided.
d. The laboratory shall adhere to the sample and extract holding times prescribed in
the analytical methods.
e. Chain-of-custody information must include:
i. Sample number;
ii. Sample description including any preservation (e.g., chemical, thermal, etc.)
used;
iii. Date and time of sample collection;
iv. Specific location of sample collection;
v. Name of sample collector and intermediate custodians, if any;
vi. Date(s) and time(s) of custody transfer to the laboratory; and
vii. Name(s) and signature(s) of the individual(s) receiving the sample.
f. A chain-of-custody form must accompany all samples including those shipped by
mail or courier.
g. The laboratory shall maintain a system of internal sample tracking that documents
sample custody from the time of receipt at the laboratory to the time of disposal.
8. Analytical Methodology. The laboratory shall utilize acceptable analytical methods.
The acceptable methods shall be those defined or referenced in 40 CFR parts 136, 141
and 143, and 310 CMR 22.00: Drinking Water, for the environmental matrix being
tested.
(b) Additional Requirements for Chemical and Radiochemical Laboratories.
1. Quality Control Procedures.
a. Unless directed otherwise by the analytical method or the Department standard
employed, the laboratory shall prepare and analyze a laboratory reagent blank, sample
duplicate, and laboratory fortified blank for every 20 or fewer samples processed as
an analytical batch. Duplicates of radiochemical samples must be prepared and
analyzed for every ten or fewer samples. In addition, a laboratory fortified sample
matrix shall be run if required by the Department. Corrective action shall be taken
if the results of these analyses do not meet acceptance criteria developed within the
laboratory according to accepted analytical procedures. The preparation of blanks,
laboratory fortified sample matrices, and duplicates and the results of their analyses
shall be recorded.
b. Each laboratory must establish acceptance limits for precision and accuracy and
maintain and use quality control charts for each of the analytes in the matrices,
disciplines, and categories in which the laboratory is certified. These limits may not
be less stringent than those defined in approved analytical methods or approved by
the Department.
c. Certified laboratories shall utilize surrogate analytes as required by the analytical
procedure employed. Acceptance limits for surrogate analyte recoveries shall be
established by the laboratory. Quality control charts must be maintained and used for
each surrogate.
d. Certified laboratories shall perform and document all quality control procedures
in established analytical protocols or the quality control procedures the Department
requires and specifies.
e. When integrating chromatography peaks, either automatically or manually, each
laboratory must ensure that integrations are performed in a correct and consistent
manner for standards and samples, including quality control samples. Each laboratory
must maintain documentation of manual integrations that includes the following:
i. The laboratory's written procedure for manual integration;
ii. The original chromatogram and the manually integrated chromatogram; and
iii. The analyst's initials, date of manual integration, and the reason(s) for the
manual integration.
2. Determination of Method Detection Limit.
a. Each laboratory shall experimentally determine the method detection limit for
analysis of each analyte, except pH, for each matrix in which the laboratory is
certified.
b. The laboratory must document its procedure for determining the method detection
limit. The laboratory must use the procedure for determining the method detection
limit that is described in the analytical method being used. If the analytical method
does not include a procedure for the determination of method detection limits, then
the laboratory must determine the method detection limit using the procedure
described in 40 CFR Part 136: Appendix B or other Department- or EPA-approved
procedure. For those analytes requiring pattern recognition for identification (e.g.,
chlordane, toxaphene), the method detection limit is defined as the lowest
concentration at which pattern recognition is possible.
c. Calculations and supporting documentation used in determining limits must be
available for inspection.
d. Detection limits shall be expressed in appropriate units.
e. The laboratory must achieve the method detection limits required by the
applicable regulations.
f. Sample preparation and analyses for the method detection limit calculation must
be made over a period of at least three days.
g. Method detection limits must be determined as part of a laboratory’s initial
demonstration of capability to perform an analysis, when any change occurs in the
laboratory that could affect the method detection limits, and as required by an
analytical method.
h. Method detection limits must be determined using analysts and instruments that
are representative of those used in the performance of analyses.
i. The laboratory must verify its capability to analyze low level samples on an
ongoing basis through the analysis of low level standards or through a method
detection limit determination.
j. The laboratory shall determine the minimum reporting level for analysis of each
analyte, except pH, for each matrix in which the laboratory is certified. The
laboratory shall document the procedure used to determine the minimum reporting
level. The laboratory shall verify the minimum reporting level on an ongoing basis.
3. Laboratory Reagent Water. The laboratory must demonstrate that its reagent water
meets the specifications required by the analytical methods it uses including that it is free
of analytes of interest above their method detection limits. When the method specifies
the resistance or conductivity of the source of reagent water, verification of such quality
shall be made and documented each day the water is used.
(c) Additional Requirements for Microbiology Laboratories.
1. Autoclaves.
a. For each sterilization cycle, the signature of the analyst, date, sterilization time
and temperature, the materials being autoclaved and their total time in the autoclave
shall be recorded.
b. A maximum-temperature-registering thermometer, electronic temperature read-
out device, or continuous recording device must be used during each autoclave cycle.
The temperature must be recorded.
c. Automatic timing mechanisms must be checked quarterly with an accurate
timepiece, such as a stopwatch, and the results recorded.
d. The laboratory must check the performance of the autoclave each week during
which the autoclave is used with Geobacillus stearothermophilus spore strips,
suspensions, or capsules and record the results.
e. Biological waste must be autoclaved for at least 30 minutes followed by proper
disposal.
2. Hot Air Sterilizing Ovens. The laboratory must check the performance of the hot air
sterilizing oven each week during which the oven is used with Bacillus subtilis or
Bacillus atrophaeus spore strips and record the results.
3. Incubators.
a. Incubators, both air-type and water bath, must maintain the temperature specified
by the method. On days when the incubator is in use, the temperature of each
incubator must be recorded continuously or at least twice per day, with each reading
separated by at least four hours and the times of each reading recorded. For air-type
incubators, the thermometer shall be immersed in liquid and placed on one of the
shelves in use.
b. If an aluminum block incubator is used, culture dishes and tubes must fit snugly.
c. Water bath incubators must be cleaned at least monthly.
d. The laboratory must record the date and time at the beginning and at the end of
sample incubation.
4. Germicidal Ultraviolet Lamps. Germicidal ultraviolet lamps shall be tested quarterly
by exposing agar spread plates containing 200 to 250 microorganisms to the light for two
minutes. If such irradiation does not reduce the count of control plates by 99%, the
lamps shall be replaced. Alternatively, the laboratory shall use an ultraviolet light meter
to ensure that the lamp emits at least 70% of its initial output. Cleaning of germicidal
ultraviolet lamps shall be done at least monthly by disconnecting the unit and cleaning
the lamps with a soft cloth moistened with ethanol.
5. Microscopes. The optics and stage of microscopes shall be cleaned with lens paper
prior to each use.
6. Sterility of Rinse/Dilution Water and Sample Bottles.
a. Each batch or lot of dilution/rinse water must be checked for sterility by adding
50 mL of the water to a 50 mL volume of a double strength non-selective broth (e.g.,
tryptic soy broth, trypticase soy broth, or tryptose broth), which is then incubated at
35°C + 0.5°C and checked for turbidity signifying growth at 24 and 48 hours and the
results recorded.
b. After sterilization, at least one bottle per batch of sterilized sample bottles or per
lot of commercially prepared sample containers shall be checked for sterility by
adding approximately 25 mL of sterile non-selective broth medium to each bottle.
The bottle shall be capped and rotated so that the broth comes in contact with all
surfaces and shall be incubated at 35°C + 0.5°C and checked for growth at 24 and 48
hours and the results recorded. Prepared sample bottles from each batch or lot shall
not be used unless satisfactory results are obtained from the tested bottle.
7. Residue Testing of Glassware.
a. Inhibitory Residue Test. With the initial use of each lot of a detergent or washing
product, the rinsing process using distilled or deionized water shall be demonstrated
to provide glassware that is free from toxic material based on the use of the Inhibitory
Residue Test, as specified in the most recent edition of Standard Methods for the
Examination of Water and Wastewater, American Public Health Association,
American Water Works Association, Water Environment Federation, Washington
D.C. The results of the test must be recorded. Alternatively, the laboratory may
obtain written, traceable certification from the product manufacturer that the
inhibitory residue test has been performed on the lot of detergent or washing product
according to the Standard Methods procedure. The actual test results must be
included with the certification.
b. Bromthymol Blue Test. Each batch of clean, dry glassware or plasticware shall
be tested for residual alkaline or acid residue using bromthymol blue indicator and
the results recorded. If the results of the indicator test are not within the desired color
range of light green to dark blue, corrective action shall be taken by re-rinsing, air
drying and retesting.
8. Microbiological Media - Quality Control Measures.
a. The laboratory shall keep records that indicate the kind, amount, date received,
lot number, expiration date, and date of opening of bottles of media. Media shall be
stored in a desiccator or cool, dry location. If caking or discoloration of media
occurs, media shall be discarded.
b. Records shall be available for inspection on all batches of laboratory-prepared
media showing lot numbers, date prepared, details of preparation, total volume
prepared, sterilization time and temperatures, final pH, and the name of the
individual who performed the work.
c. Prior to first use of media, the laboratory shall test each batch of medium
prepared in the laboratory and each lot of pre-prepared, ready-to-use medium with
at least one pure culture of a known positive reaction.
d. Prior to first use of media, the laboratory shall test each batch of medium prepared
in the laboratory and each lot of pre-prepared, ready-to-use medium with one or more
negative culture controls, i.e., non-target organisms, as appropriate to the method.
e. Prepared plates shall be refrigerated in sealed containers with a label containing
the date of preparation or expiration and the name of the medium. Plates may be kept
no more than two weeks following preparation. Broth media in loose-capped test
tubes may be kept no more than two weeks following preparation. Broth media in
tightly capped tubes may be kept three months from the date of preparation.
9. Dechlorination Sufficiency. If chlorinated water is to be analyzed, sufficient sodium
thiosulfate must be added to the sample bottle before sterilization to neutralize any
residual chlorine in the water sample. The laboratory may also use commercially
prepared, pre-sterilized bags or bottles containing sodium thiosulfate.
10. Membrane Filter Procedure Quality Control Specifics.
a. Only membrane filters recommended for water analysis by the manufacturer shall
be used.
b. Lot numbers of membrane filters and date of receipt shall be recorded.
c. Procedural Contamination. A start and finish membrane filtration control test of
rinse water, medium, and supplies shall be conducted for each filtration series. If
sterile controls indicate contamination, all data on samples affected shall be rejected
and a request made for immediate resampling of those waters affected.
d. Verification of Membrane Filter Colonies on m-Endo medium.
Total Coliform Procedure. All sheen or borderline colonies up to ten on each
membrane shall be verified in accordance with the accepted standard procedure
contained in the latest edition of Standard Methods for the Examination of Water and
Wastewater (Standard Methods).
11. Quality Control.
a. When quality control samples are available, each analyst shall analyze at least one
quality control sample per year for the categories to be certified.
b. During the initial training of an analyst in a method requiring the identification
and enumeration of colonies, the new analyst must count plates from at least ten
positive samples having varying colony counts within the ideal counting range for the
method. The laboratory supervisor must count the same plates. The replicate counts
between the analysts must agree within 10%.
c. For methods used for the enumeration of colonies, 10% of routine samples must
be analyzed in duplicate and the range of logs determined. Corrective action shall
be taken if the results of these analyses do not meet acceptance criteria developed
within the laboratory according to accepted analytical procedures.
12. Laboratory Reagent Water. The laboratory shall use satisfactorily tested reagent
water from a water purification system (e.g., still, deionization unit, or a reverse-osmosis
unit) to prepare media, reagents, and dilution/rinse water for performing microbial
analyses.
a. If the source water is chlorinated, the laboratory reagent water must be tested
monthly for total chlorine residual and found to contain less than 0.1 mg/L of
chlorine residual.
b. The laboratory reagent water must be analyzed annually for the presence of lead,
cadmium, chromium, copper, nickel, and zinc. The concentration of each metal must
be no greater than 0.05 mg/L. The concentration of the metals collectively must be
no greater than 0.1 mg/L.
c. The conductivity of the laboratory reagent water must be monitored each day that
the water is used and found to be less than 2 :S/cm at 25/C or greater than 0.5
megohms-cm resistance at 25/C.
d. A heterotrophic plate count must be performed on the laboratory reagent water
monthly. The laboratory may use the reagent water only if the heterotrophic plate
count is less than 500 CFU/mL.
e. The test of the bacteriological quality of the laboratory reagent water must be
performed annually. The ratio of the growth rate must be between 0.8 and 3.0. The
test is described in Section 9020B of the 18 and 19 editions of Standard Methods
th
th
for the Examination of Water and Wastewater. The bacteriological quality test is not
required of laboratories that document that their laboratory reagent water meets the
criteria for Type II water or better, as defined in Standard Methods (18 and 19
th
th
editions), Section 1080C or Medium quality water or better as defined in Standard
Methods (20 edition), Section 1080C.
th
(6) On an annual basis, certified laboratories and laboratories seeking certification shall require
all personnel to participate in a laboratory ethics training program. Training shall include the
following topics: proper procedures to ensure data integrity, recognition and prevention of
improper laboratory practices, the promotion of objectivity and impartiality in the generation and
reporting of analytical data, and procedures for confidential reporting of data integrity concerns
to a laboratory supervisor, director, or owner, as appropriate. The laboratory must document the
content of the training and the date of participation in the training for each staff member and
shall make this documentation and the materials used in the training available for review during
an inspection.