105 CMR 120.225
General
(A) Each licensee or registrant shall make, or cause to be made, surveys of areas, including the
subsurface, that:
(1) are necessary for the licensee or registrant to comply with 105 CMR 120.200; and
(2) are necessary under the circumstances to evaluate:
(a) the magnitude and extent of radiation levels;
(b) concentrations or quantities of radioactive material residual radioactivity; and
(c) the potential radiological hazards of the radiation levels and residual radioactivity
detected.
(B) Notwithstanding the provisions in 105 CMR 120.263(A), records from surveys describing
the location and amount of subsurface residual radioactivity identified at the site must be kept
with records important for decommissioning, and such records must be retained in accordance
with 105 CMR 120.125(C)(8), as applicable.
(C) The licensee or registrant shall ensure that instruments and equipment used for quantitative
radiation measurements, for example, dose rate and effluent monitoring, are calibrated at
intervals not to exceed 12 months for the radiation measured except when a more frequent
interval is specified in another applicable section of 105 CMR 120.000 or license condition.
(D) All personnel dosimeters, except for direct and indirect reading pocket ionization chambers
and those dosimeters used to measure the dose to any extremity, that require processing to
determine the radiation dose and that are used by licensees and registrants to comply with
105 CMR 120.211, with other applicable provisions of 105 CMR 120.000, or with conditions
specified in a license or certificate of registration, shall be processed and evaluated by a
dosimetry processor:
(1) holding current personnel dosimetry accreditation from the National Voluntary
Laboratory Accreditation Program (NVLAP) of the National Institute of Standards and
Technology; and
(2) approved in this accreditation process for the type of radiation or radiations included in
the NVLAP program that most closely approximates the type of radiation or radiations for
which the individual wearing the dosimeter is monitored.
(E) The licensee or registrant shall ensure that adequate precautions are taken to prevent a
deceptive exposure of an individual monitoring device.