105 CMR 120.314
Radiation Survey Instruments
(A) The licensee or registrant shall maintain sufficient calibrated and operable radiation survey
instruments to make physical radiation surveys as required by 105 CMR 120.300 and
120.225(A). Instrumentation required by 105 CMR 120.300 shall have a range from 0.02
mSv/hr (2 mrem/hr) through 0.01 Sv/hr (1 rem/hr).
(B) Each radiation survey instrument shall be calibrated:
(1) By a person licensed or registered by the Agency, another Agreement State, a Licensing
State, or the U.S. Nuclear Regulatory Commission to perform such service;
(2) At energies appropriate for the licensee's or registrant's use;
(3) At intervals not to exceed six months and after each instrument servicing other than
battery replacement;
(4) To demonstrate an accuracy within plus or minus 20%; and
(5) At two points located approximately a and b of full-scale on each scale for linear scale
instruments; at midrange of each decade, and at two points of at least one decade for
logarithmic scale instruments; and for digital instruments, at three points between 0.02 and
10 mSv/hr (2 and 1,000 mrem/hr).
(C) Records of these calibrations shall be maintained for Agency inspection for five years after
the calibration date.
(D) Each radiation survey instrument shall be checked with a radiation source at the beginning
of each day of use and at the beginning of each work shift to ensure it is operating properly.