105 CMR 120.420
Appendix A -- Radiation Shielding and Safety Requirements
In order for the Radiation Control Program to provide an evaluation, and official approval
on shielding requirements for a radiation installation, the following must be submitted. The
plans shall show as a minimum the following:
(A) The normal location of the x-ray system's radiation port, the port's travel and transverse
limits, general direction(s) of the useful beam; locations of any windows and doors; the location
of the operator's booth and the location of the x-ray control panel.
(B) The structural composition and thickness or lead equivalent of all walls, doors, partitions,
floor and ceiling of the room(s) concerned.
(C) The dimensions of the room(s) concerned.
(D) The type of occupancy of all adjacent areas inclusive of space above and below the room(s)
concerned. If there is any exterior wall, show distance to the closest area(s) where it is likely that
individuals will be present.
(E) The make and model of the x-ray equipment.
(F) The typical type of examination(s) and treatment(s) which will be performed with the
equipment.
(G) Information on the anticipated workload of the x-ray system(s).
(H) An interlock and/or warning light shall be installed at all egresses. For diagnostic x-ray
installations, the warning light shall be wired to the rotor of the x-ray system.
(I) All basic assumptions used to determine the shielding requirements in developing these
plans shall be submitted with these plans.