105 CMR 120.436
Therapeutic Radiation Machines of Less than 500 kV
(A) Leakage Radiation. When the x-ray tube is operated at its maximum rated tube current for
the maximum kV, the leakage air kerma rate shall not exceed the value specified at the distance
specified for that classification of therapeutic radiation machine:
(1) 5-50 kV Systems. The leakage air kerma rate measured at any position five centimeters
from the tube housing assembly shall not exceed one mGy (100 mrad) in any one hour.
(2) >50 and <500 kV Systems. The leakage air kerma rate measured at a distance of one
meter from the target in any direction shall not exceed one cGy (one rad) in any one hour.
This air kerma rate measurement may be averaged over areas no larger than 100 square
centimeters. In addition, the air kerma rate at a distance of five centimeters from the surface
of the tube housing assembly shall not exceed 30 cGy (30 rad) per hour.
(3) For each therapeutic radiation machine, the registrant shall determine, or obtain from
the manufacturer, the leakage radiation existing at the positions specified in 105 CMR
120.436(A)(1) and (2) for the specified operating conditions. Records on leakage radiation
measurements shall be maintained at the installation for inspection by the Agency.
(B) Permanent Beam Limiting Devices. Permanent diaphragms or cones used for limiting the
useful beam shall provide at least the same degree of attenuation as required for the tube housing
assembly.
(C) Adjustable or Removable Beam Limiting Devices.
(1) All adjustable or removable beam limiting devices, diaphragms, cones or blocks shall
not transmit more than 5% of the useful beam for the most penetrating beam used;
(2) When adjustable beam limiting devices are used, the position and shape of the radiation
field shall be indicated by a light beam.
(D) Filter System. The filter system shall be so designed that:
(1) Filters can not be accidentally displaced at any possible tube orientation;
(2) For equipment installed after July 9, 1999, an interlock system prevents irradiation if the
proper filter is not in place;
(3) The air kerma rate at a distance of one meter from the filter shall not exceed one cGy
(one rad) per hour under any operating conditions; and,
(4) Each filter shall be marked as to its material of construction and its thickness.
(E) Tube Immobilization.
(1) The x-ray tube shall be so mounted that it can not accidentally turn or slide with respect
to the housing aperture; and
(2) The tube housing assembly shall be capable of being immobilized for stationary portal
treatments.
(F) Source Marking. The tube housing assembly shall be so marked that it is possible to
determine the location of the source to within five millimeters, and such marking shall be readily
accessible for use during calibration procedures.
(G) Beam Block. Contact therapy tube housing assemblies shall have a removable shield of
material, equivalent in attenuation to 0.5 millimeters of lead at 100 kV, which can be positioned
over the entire useful beam exit port during periods when the beam is not in use.
(H) Timer. A suitable irradiation control device shall be provided to terminate the irradiation
after a pre-set time interval.
(1) A timer which has a display shall be provided at the treatment control panel. The timer
shall have a pre-set time selector and an elapsed time or time remaining indicator;
(2) The timer shall be a cumulative timer which activates with an indication of
"BEAM-ON" and retains its reading after irradiation is interrupted or terminated. After
irradiation is terminated and before irradiation can be reinitiated, it shall be necessary to reset
the elapsed time indicator;
(3) The timer shall terminate irradiation when a pre-selected time has elapsed, if any dose
monitoring system present has not previously terminated irradiation;
(4) The timer shall permit accurate pre-setting and determination of exposure times as short
as one second;
(5) The timer shall not permit an exposure if set at zero;
(6) The timer shall not activate until the shutter is opened when irradiation is controlled by
a shutter mechanism unless calibration includes a timer error correction to compensate for
mechanical lag; and,
(7) The timer and back-up timer if present shall be accurate to within 1% of the selected
value or one second, whichever is greater.
(I) Control Panel Functions. The control panel, in addition to the displays required by other
provisions in 105 CMR 120.436, shall have:
(1) An indication of whether electrical power is available at the control panel and if
activation of the x-ray tube is possible;
(2) An indication of whether x-rays are being produced;
(3) Means for indicating x-ray tube potential and current;
(4) The means for terminating an exposure at any time;
(5) A locking device which will prevent unauthorized use of the therapeutic radiation
machine; and
(6) For therapeutic radiation machines manufactured after July 9, 1999, a positive display
of specific filter(s) in the beam.
(J) Multiple Tubes. When a control panel is capable of energizing more than one x-ray tube:
(1) It shall only be allowable to activate one X-ray tube at any time;
(2) There shall be an indication at the control panel identifying which x-ray tube is
activated; and
(3) There shall be an indication at the tube housing assembly when that tube is energized.
(K) Target-to-skin Distance (TSD). There shall be a means of determining the central axis TSD
to within one centimeter and of reproducing this measurement to within two millimeters
thereafter.
(L) Shutters. Unless it is possible to bring the x-ray output to the prescribed exposure
parameters within five seconds after the x-ray "ON" switch is energized, the beam shall be
attenuated by a shutter having a lead equivalency not less than that of the tube housing assembly.
In addition, after the unit is at operating parameters, the shutter shall be controlled by the
operator from the control panel. An indication of shutter position shall appear at the control
panel.
(M) Low-filtration X-ray Tubes. Each therapeutic radiation machine equipped with a beryllium
or other low-filtration window shall be clearly labeled as such upon the tube housing assembly
and shall be provided with a permanent warning device on the control panel that is activated
when no additional filtration is present, to indicate that the dose rate is very high.
(N) Facility Design Requirements for Therapeutic Radiation Machines Capable of Operating
in the Range 50 kV to 500 kV. In addition to shielding adequate to meet requirements of
105 CMR 120.439, the treatment room shall meet the following design requirements:
(1) Aural Communication. Provision shall be made for continuous two-way aural
communication between the patient and the operator at the control panel;
(2) Viewing Systems. Provision shall be made to permit continuous observation of the
patient during irradiation and the viewing system shall be so located that the operator can
observe the patient from the control panel. The therapeutic radiation machine shall not be
used for patient irradiation unless at least one viewing system is operational.
(O) Additional Requirements. Treatment rooms which contain a therapeutic radiation machine
capable of operating above 150 kV shall meet the following additional requirements:
(1) All protective barriers shall be fixed except for entrance doors or beam interceptors;
(2) The control panel shall be located outside the treatment room or inside the treatment
room within a totally enclosed booth with protective barrier walls, door(s), ceiling and floor;
(3) Interlocks shall be provided such that all entrance doors, including doors to any interior
booths, shall be closed before treatment can be initiated or continued. If the radiation beam
is interrupted by any door opening, it shall not be possible to restore the machine to operation
without closing the door and reinitiating irradiation by manual action at the control panel;
and
(4) When any door referred to in 105 CMR 120.436(O)(3) is opened while the x-ray tube
is activated, the air kerma rate at a distance of one meter from the source shall be reduced to
less than 1 mGy (100 mrad) per hour.
(P) Full Calibration Measurements.
(1) Full calibration of a therapeutic radiation machine subject to 105 CMR 120.436 shall
be performed by, or under the direct supervision of, a Qualified Medical Physicist:
(a) Before the first medical use following installation or reinstallation of the therapeutic
radiation machine;
(b) At intervals not exceeding one year; and
(c) Before medical use under the following conditions:
1. Whenever quality assurance check measurements indicate that the radiation
output differs by more than 5% from the value obtained at the last full calibration and
the difference cannot be reconciled; and
2. Following any component replacement, major repair, or modification of
components that could significantly affect the characteristics of the radiation beam.
(d) Notwithstanding the requirements of 105 CMR 120.436(P)(1)(c):
1. Full calibration of therapeutic radiation machines with multi-energy capabilities
is required only for those modes and/or energies that are not within their acceptable
range; and
2. If the repair, replacement or modification does not affect all energies, full
calibration shall be performed on the affected energy that is in most frequent clinical
use at the facility. The remaining energies may be validated with quality assurance
check procedures against the criteria in 105 CMR 120.436(P)(1)(c)1.
(2) To satisfy the requirement of 105 CMR 120.436(P)(1), full calibration shall include all
measurements recommended for annual calibration by NCRP Report 69, "Dosimetry of
X-ray and Gamma Ray Beams for Radiation Therapy in the Energy Range 10 keV to 50
MeV" (1981).
(3) The registrant shall maintain a record of each calibration for the duration of the
registration. The record shall include the date of the calibration, the manufacturer's name,
model number, and serial number for both the therapeutic radiation machine and the X-ray
tube, the model numbers and serial numbers of the instruments used to calibrate the
therapeutic radiation machine, and the signature of the Qualified Medical Physicist
responsible for performing the calibration.
(Q) Periodic Quality Assurance Checks.
(1) Periodic quality assurance checks shall be performed on therapeutic radiation machines
subject to 105 CMR 120.436, which are capable of operation at greater than or equal to 50
kV.
(2) To satisfy the requirement of 105 CMR 120.436(Q)(1), quality assurance checks shall
meet the following requirements:
(a) The registrant shall perform quality assurance checks in accordance with written
procedures established by the Qualified Medical Physicist; and
(b) The quality assurance check procedures shall specify the frequency at which tests
or measurements are to be performed. The quality assurance check procedures shall
specify that the quality assurance check shall be performed during the calibration
specified in 105 CMR 120.436(P)(1). They shall also state the acceptable tolerance for
each parameter measured in the quality assurance check, when compared to the value for
that parameter determined in the calibration specified in 105 CMR 120.436(P)(1), shall
be stated.
(3) The cause for a parameter exceeding a tolerance set by the Qualified Medical Physicist
shall be investigated and corrected before the system is used for patient irradiation;
(4) Whenever a quality assurance check indicates a significant change in the operating
characteristics of a system, as specified in the Qualified Medical Physicist's quality assurance
check procedures, the system shall be recalibrated as required in 105 CMR 120.436(P)(1);
(5) The registrant shall use the dosimetry system described in 105 CMR 120.434(C)(2) to
make the quality assurance check required in 105 CMR 120.436(Q)(2);
(6) The registrant shall have the Qualified Medical Physicist review and sign the results of
each radiation output quality assurance check within 30 days of the date that the check was
performed;
(7) The registrant shall ensure that safety quality assurance checks of therapeutic radiation
machines subject to 105 CMR 120.436 are performed at intervals not to exceed 30 days;
(8) Notwithstanding the requirements of 105 CMR 120.436(Q)(6) and (7), the registrant
shall ensure that no therapeutic radiation machine is used to administer radiation to humans
unless the quality assurance checks required by 105 CMR 120.436(Q)(6) and (7) have been
performed within one month period immediately prior to said administration;
(9) To satisfy the requirement of 105 CMR 120.436(Q)(7), safety quality assurance checks
shall ensure proper operation of:
(a) Electrical interlocks at each external beam radiation therapy room entrance;
(b) Proper operation of the "BEAM-ON" and termination switches;
(c) Beam condition indicator lights on the access door(s), control console, and in the
radiation therapy room;
(d) Viewing systems;
(e) If applicable, electrically operated treatment room doors activated from inside and
outside the treatment room;
(10) The registrant shall maintain a record of each quality assurance check required by
105 CMR 120.436(Q)(1) and (7) for three years. The record shall include the date of the
quality assurance check, the manufacturer's name, model number, and serial number for the
therapeutic radiation machine, the manufacturer's name, model number and serial number
of the instrument(s) used to measure the radiation output of the therapeutic radiation
machine, and the signature of the individual who performed the periodic quality assurance
check.
(R) Operating Procedures.
(1) The therapeutic radiation machine shall not be used for irradiation of patients unless and
until the requirements of 105 CMR 120.436(P) and (Q) have been met;
(2) Therapeutic radiation machines shall not be left unattended unless secured pursuant to
105 CMR 120.436(I)(5);
(3) Mechanical supporting or restraining devices shall be used when a patient must be held
in position for radiation therapy;
(4) The tube housing assembly shall not be held by an individual during operation unless
the assembly is designed to require such holding and the peak tube potential of the system
does not exceed 50 kV. In such cases, the holder shall wear protective gloves and apron of
not less than 0.5 millimeters lead equivalency at 100 kV;
(5) A copy of the current operating and emergency procedures shall be maintained at the
therapeutic radiation machine control console; and
(6) No individual other than the patient shall be in the treatment room during exposures
from therapeutic radiation machines operating above 150 kV. At energies less than or equal
to 150 kV, any individual, other than the patient, in the treatment room shall be protected by
a barrier sufficient to meet the requirements of 105 CMR 120.211.
(S) Possession of Survey Instrument(s). Each facility location authorized to use a therapeutic
radiation machine in accordance with 105 CMR 120.436 shall possess appropriately calibrated
portable monitoring equipment. As a minimum, such equipment shall include a portable
radiation measurement survey instrument capable of measuring dose rates over the range ten µSv
(one mrem) per hour to ten mSv (1000 mrem) per hour. The survey instrument(s) shall be
operable and calibrated in accordance with 105 CMR 120.438.