105 CMR 120.438
Calibration of Survey Instruments
(A) The registrant shall ensure that the survey instruments used to show compliance with
105 CMR 120.430 have been calibrated before first use, at intervals not to exceed 12 months,
and following repair.
(B) To satisfy the requirements of 105 CMR 120.438(A), the registrant shall:
(1) Calibrate all required scale readings up to ten mSv (1000 mrem) per hour with an
appropriate radiation source that is traceable to the National Institute of Standards and
Technology (NIST);
(2) Calibrate at least two points on each scale to be calibrated. These points should be at
approximately a and b of full-scale; and
(3) Calibrate automatically ranging digital display survey instruments at no less than one
point on each decade and at no less than two points on one of these decades. These points
should be at approximately a and b of the decade.
(C) To satisfy the requirements of 105 CMR 120.438(B), the registrant shall:
(1) Consider a point as calibrated if the indicated dose rate differs from the calculated dose
rate by not more than 10%; and,
(2) Consider a point as calibrated if the indicated dose rate differs from the calculated dose
rate by not more than 20% if a correction factor or graph is conspicuously attached to the
instrument.
(D) The registrant shall retain a record of each calibration required in 105 CMR 120.438(A) for
three years. The record shall include:
(1) A description of the calibration procedure; and
(2) A description of the source used and the certified dose rates from the source, and the
rates indicated by the instrument being calibrated, the correction factors deduced from the
calibration data, the signature of the individual who performed the calibration, and the date
of calibration.
(E) The registrant may obtain the services of individuals licensed by the Agency, the U.S.
Nuclear Regulatory Commission, an Agreement State, or a Licensing State to perform
calibrations of survey instruments. Records of calibrations which contain information required
by 105 CMR 120.438(D) shall be maintained by the registrant.