Authorization for Calibration, Transmission and Reference Sources

Year: 2026Length: 176 wordsOfficial source
(A) Any person authorized by 105 CMR 120.506 for medical use of byproduct material may receive, possess, and use any of the following byproduct material for check, calibration, transmission, and reference use: (1) Sealed sources, not exceeding 1.11 GBq (30 mCi) each, manufactured and distributed by a person licensed under 105 CMR 120.128(L) or equivalent Nuclear Regulatory Commission or Agreement State regulations; (2) Sealed sources, not exceeding 1.11 Gbq (30 mCi) each, redistributed by a licensee authorized to redistribute the sealed sources manufactured and distributed by a person licensed under 105 CMR 120.128(L) or equivalent Nuclear Regulatory Commission or Agreement State regulations, providing the redistributed sealed sources are in the original packaging and shielding and are accompanied by the manufacturer's approved instructions; (3) Any byproduct material with a half-life not longer than 120 days in individual amounts not to exceed 0.56 GBq (15 mCi); (4) Any byproduct material with a half-life longer than 120 days in individual amounts not to exceed the smaller of 7.4 MBq (200 µCi) or 1000 times the quantities in 105 CMR